Austin Scientifi E1000 Manual do utilizador - Página 6
Procurar online ou descarregar pdf Manual do utilizador para Controlador de temperatura Austin Scientifi E1000. Austin Scientifi E1000 18 páginas. Dual temperature controller
6 – Mode Selection. 2 miniature slide
switches allow the user to select either
Coldhead or Waterpump mode. In the
coldhead mode, the channel will provide
power to the output at all times and will
operate the relays according to the user
set temperatures. In the Cryopump mode,
power will be supplied at all times to the
coldhead, unless, the manual mode is
selected. To change the Mode of the
E1000, the power switch must be turned
off and back on after changing the
positioni of the switch.
3
•
•
•
/0111 %
/0111 %
/0111 %
/0111 %
2
2
2
2
7 – Compressor Power in.
power from the compressor. Requires
special drive cable.
8 – Compressor Power Out. Power out to
drive and additional E1000. The power is
fused at 3 amps.
9 – Coldhead out - Output power to the
coldhead or water pump. Uses standard
cryopump drive cable.
The Business of Science
Input drive
®