Great Lakes MINISTAR S Manual de operação - Página 6
Procurar online ou descarregar pdf Manual de operação para Equipamento dentário Great Lakes MINISTAR S. Great Lakes MINISTAR S 13 páginas.
Também para Great Lakes MINISTAR S: Dicas de resolução de problemas e fabrico (2 páginas)
Maintenance
A regular maintenance schedule should be followed. Proper maintenance will assure better fabrication results. Refrain
from using water in the MiniSTAR S machine. Water may cause certain components to rust or corrode. When the machine
is not being used, engage the clamping frame to the chamber and position the chamber on the pellet cup. This prevents
lab dust from entering the pellet cup and chamber.
D
D
Pellets will become dirty with regular use. Clean pellets by
placing them in a strainer and rinsing with hot water. Spread
pellets on a towel and let air dry before placing them back in
the MiniSTAR S. Check pellets monthly. Discard and replace
pellets in the bottom third of the cup.
Air escape holes are present in the base of the pellet cup and
should be checked monthly to ensure that they remain open.
Use a .036 " diameter or smaller wire to open plugged holes.
Clean and inspect top and side of the Adjusting
every 4-6 months.
• Clean gasket with a soft cloth
A
• Check potential gasket cuts or irregularities
*Adjusting Ring is assembled from four components:
A.
Chamber sealing ring
B.
Aluminum ring
C.
Side gasket
D
D.
Spring-loaded adjusting pins
B
*Adjusting Ring should move freely in the chamber. If not,
C
after cleaning, apply a thin coat of Silicone Grease (016-021)
to inner wall of the chamber, 4 spring-loaded adjusting pins,
D
and outside edge of the adjusting ring (see white arrows).
Ring*
6