STEP 1 – UNPACKING THE AFFLOVEST
Carefully unpack the AffloVest and its accessories.
You must verify that the battery is fully charged before using the vest.
STEP 2 - CHECKING BATTERY LEVEL AND CHARGING BATTERY
IMPORTANT
Use only the supplied rechargeable battery and the supplied power supply to ensure
safe and reliable operation of the AffloVest.
The battery must be fully charged before using the AffloVest for the first time.
To check the battery level, depress the fuel gauge push-button to activate the fuel
gauge LED's.
The battery LED's will light up to indicate the battery level.
1 LED blinking
2 LED's lit
(0 - 19%)
(20 - 39%)
To charge the battery, connect the power supply to a wall outlet and plug the power
supply into the battery port as shown. The power supply LED will turn on when
connected to the wall outlet.
10
Battery level push button
3 LED's lit
4 LED's lit
5 LED's lit
(40 - 59%)
(60 - 79%)
(80 - 100%)
Power Supply LED lit
Next, plug the power supply plug into the AffloVest battery. There is a small opening
at the bottom of the battery pouch.
Power supply connected to battery
IMPORTANT
• When the power supply is plugged into the battery, pay close attention not to
place any strain onto the power supply cord. Rough handling of the vest and
power supply while plugged into the battery could cause damage to the battery.
• When the battery is charging, the LED's will become lit with one (1) LED blinking.
Note, this may take 5 – 10 seconds for the LED's to light after plugging in the
power supply.
• It is recommended to allow the battery fully charge for at least three (3) hours and
until all LED's are solidly lit. At the completion of charging, disconnect the power
supply from the battery. It is recommended that the power supply be unplugged
from the wall outlet and stored until next use.
• The battery gauge LED's will remain lit for approximately one (1) minute after
removing the power supply.
11