Gait up Physilog 5 Manual do utilizador - Página 3
Procurar online ou descarregar pdf Manual do utilizador para Sensores de segurança Gait up Physilog 5. Gait up Physilog 5 17 páginas.
Também para Gait up Physilog 5: Manual do utilizador (12 páginas), Manual do utilizador (17 páginas)
Supply Voltage specifications
The Physilog® 5 should be charged using USB at 5V, 100mA. For charging, the ambient temperature
must lie between 0°C and 45°C.
Recording with Physilog®5
3.
There are two ways to start and stop a measurement with the Physilog®5:
3.1. Standalone measurement
Press on the main button (for about 2 secs, until the green light turns off) to start the measurement.
During measurement, the Physilog® blinks green.
Hold the main button for 3 seconds, until the light turns orange to stop the measurement. When
the measurement has been concluded, the LED shortly blinks green three times.
3.2. Standalone synchronous measurement with multiple Physilog®5
Starting from firmware version 1.1.0 (June 2017):
The Physilog®5 sensors synchronize regularly during the measurement. A network includes one
single master sensor and several slave sensors (from one to as many as wished). Multiple sensor
networks can be created by selecting a different radio channel. The network configuration is done
via the Research ToolKit Desktop software (see RTK user manual for details) and is saved in the
conf.bin file on each Physilog®5.
To perform a measurement with synchronized Physilog®5, turn on all sensors of the network with
the main button (like for standalone measurement above) with short time intervals (all must be
started within one minute). No precise order of starting the sensors is necessary. Each sensor starts
recording data from start and once it is synchronized with the master sensor it will blink green
synchronously with the master sensor. To stop the measurement, push the main button of the
sensors until the LED turns orange, you can turn off the sensors in any order.
Physilog5 user manual v1.2.6 Date: 19.06.2018
Copyright © 2018, All Rights Reserved - Gait Up SA. CH-1015 Switzerland
EPFL Innovation Park, Bat C. CH-1015 Lausanne
TVA CHE-145.136.760
Gait Up SA
www.gaitup.com
3