• Open the accessory box and use this list to make sure you received all
components of the Vet-Scan HM5 system.
• VetScan HM5 Analyzer.
• VetScan HM5 Operator's Manual.
• External power supply.
• Mini-keyboard
• Four sample tube adaptors.
• Reagent Tubing Kit.
Selecting a Location
The analyzer must be installed in a suitable location.
• The location should be flat, level, sturdy, vibration-free, and as dust-free as possible.
• Leave at least 50 cm of clearance on all sides of the instrument to allow for
adequate airflow and access to connectors.
• Maintain a minimum of 20 cm behind the analyzer's rear panel for heat dissipation
and tube clearance.
• Leave enough space for the reagent pack.
Connecting the Reagent Pack
• Place the reagent pack near the analyzer, either beside
or behind the instrument.
• For best results, place the reagent pack at the same level as
the analyzer. If you have to place the reagent pack on a lower
level, make sure that it is no more than 45 cm below the
level of the analyzer's reagent inlets, as measured from the
bottom of the reagent pack to the bottom of the analyzer.
Note:
Do not place the reagent pack on a higher level.
Ideal
Up to 45 cm
below is ok
< 45 cm not preferred