ABB Endura AZ40 Instruções de funcionamento - Página 9
Procurar online ou descarregar pdf Instruções de funcionamento para Equipamento industrial ABB Endura AZ40. ABB Endura AZ40 16 páginas. Oxygen and carbon monoxide equivalent (coe) analyzer
Também para ABB Endura AZ40: Manual (16 páginas), Manual de Instruções (20 páginas)
8.3 Sensor assembly gaskets
8.3.1 Removing the O
sensor
2
IMPORTANT NOTE
The O
sensor must be removed to enable access to the
2
COe block thermocouple – see Fig. 8.4, page 11, item
L.
Referring to Fig. 8.3:
1. Place the sensor assembly on a clean flat surface with the
flange studs pointing down.
2. Disconnect O
sensor wiring at upper terminals 8, 9, 10
2
using a small flat-bladed screwdriver to depress the spring
connector in each terminal and withdraw the wires from the
terminal housing.
sensor A from manifold block B using a
3. Unscrew O
2
7
/
in. spanner (wrench) and withdraw the sensor.
8
CAUTION – Minor injuries
Do not overtighten when refitting the O
1
2
3
4
8
O2 HTR WHT+
Fig. 8.3 Removing the O
sensor
2
sensor.
2
5
6
7
8
8
9 10
9 10 11 12 13 14 15 16 17 18 19 20
10
9
O2 SEN BLK
O2 HTR WHT–
Endura AZ40 | Sensor assembly | Orifice and seals kit | INS/ANAINST/007-EN Rev. B
9