The correct use of wEc1 and wTn1 units is based on the assumption that one of them must be set as
Sensor1 and the other as Sensor2: wEc1 is always set as Sensor1 and cannot be modified while wTn1 can
be set in the two different ways: pls refer to the software online Help for the operating instructions.
Allowed Combinations:
wEc1 - Sensor1 (1 beep at switch on) + wTn1 set as Sensor 2 (2 beeps at switch on). These are the factory
settings for the WPP001-ET system and for the wTn1 unit with black ring of the WPP001-ETT system:
[
]
✔
wTn1 set as Sensor 1 (1 beep at switch on) + wTn1 set as Sensor 2 (2 beeps at switch on). These are the
factory settings for the wTn1 units respectively with red and black ring of the WPP001-ETT system:
[
]
✔
Wrong Combinations:
wEc1 - Sensor 1 (1 beep at switch on) + wTn1 set as Sensor 1 (1 beep at switch on):
[ ✘]
Both the wTn1 modules set as Sensor 1 (1 beep at switch on) or set as Sensor 2 (2 beeps at switch on):
[
]
✘
[ ✘]
+
+
+
+
+