Olympus DP30BW Manual de instruções - Página 11
Procurar online ou descarregar pdf Manual de instruções para Câmara digital Olympus DP30BW. Olympus DP30BW 31 páginas. Microscope digital camera
²
@
Fig. 3
2
Installing the Camera Head
}This section describes the procedure using the combination of the
U-TV1X-2 TV adapter + U-CMAD3 C-mount camera adapter.
1. Screw in the U-CMAD3 C-mount camera adapter @ into the C-mount
thread on the bottom of the camera head ².
# The shutter ³ is a precision component that should never be touched
by the user.
Then, loosen the clamping screw | at the bottom of the U-TV1X-2 TV
adapter, screw in the U-TV1X-2 into the C-mount camera adapter and
tighten the clamping screw.
· As the photographical field is as shown below, use a TV adapter having
magnification of 0.5X to 1X (if a TV adapter with a magnification below
0.5X is used, the peripheral part of the photographed image will be
obscured.)
³
|
# If the orientation of image observed through eyepieces differ from
that of image displayed on the monitor, align them by rotating the
camera mounting orientation with respect to the microscope or
performing image rotation using the camera application software.
· If a commercially available C-mount camera adapter is used, the optical
performance of the system may not be manifested fully.
Be careful in a commercially available C-mount camera adapter or
C-mount lens
Otherwise, the threaded section will hit the inside of the camera
head and cause damage to it.
Field number
22
having a longer thread length
DP30BW
(Figs. 3 to 5)
0.5X (FN 22)
1X (FN 11)
than 4.5 mm.
8