Figure 58 – Reinstall the plastic friction lock in bore
Use a 3.0mm Allen wrench or driver to loosely reinstall
the M4x12 hex socket-head cap screw to secure the
plastic friction lock into the main housing (see
59).
Figure 59 – Reinstall the friction-lock screw
Remove the Dovetail Blocks
Use a 2.5mm Allen wrench or driver to remove the
three M3x6 hex socket-head cap screws securing the
right-hand dovetail block to the main housing (see
Figure
60).
Figure 60 – Remove screws from right dovetail block
With the three screws removed, the only thing now
holding the right-hand dovetail block onto the main
Complete Teardown, Cleaning/Lubrication, and Reassembly of the Olympus BH2-CH Substage Assembly
housing are two M2x8 alignment pins. Remove the
right-hand dovetail block (with the two alignment pins)
by pulling it straight out, away from the main housing
(see
Figure 61
pins back into the dovetail block if necessary.
Figure
Figure 61 – Remove right dovetail block from the housing
Figure 62 – Dovetail blocks with alignment pins
Use a 2.5mm Allen wrench or driver to remove the
three M3x6 hex socket-head cap screws securing the
left-hand dovetail block to the main housing (see
63).
Figure 63 – Remove screws from left dovetail block
With the three screws removed, the only thing now
holding the left-hand dovetail block onto the main
housing are two M2x8 alignment pins. Remove the left-
hand dovetail block (with the two alignment pins) by
and
Figure
62). Reinsert the alignment
Revision 1
Figure
Page 17 of 26