IMPORTANT
The BXFM focusing unit can be used as the basis for installing modules such as reflected light illumina-
tion equipment, objectives, eyepieces and observation tube for construction of a microscope system.
The BXFM can be installed on the exclusive stand or on your microscope.
SAFETY PRECAUTIONS
Fig. 1
Fig. 2
1
1. Install the microscope on a sturdy, level table or bench so as not to
block the air vents on the underside of the base.
Back
2. When combing the stand (U-ST), install the BXFM so that it forms an
angle of ±30° with respect to the pillar as shown in Fig. 1.
The compact brightfield illuminator (U-KMAS) should be installed so that,
when it is seen from the front, the lamp housing is located on its right
side.
3. When carrying the BXFM, hold the stand by the bottom with one hand
as shown in Fig. 2, and hold the focusing block with the other hand.
Front
# Do not hold the BXFM by the coarse/fine adjustment knobs or lamp
housing. Otherwise, the microscope may be damaged.
4. The surface of the lamp housing @ on the rear of the microscope be-
comes very hot. When installing the microscope system, leave spaces
of more than 10 cm around the lamp housing, particularly on the top
and bottom.
Do not install the lamp housing up side down or with the side facing up.
Also do not use the microscope in a tilted attitude.
5. The TH4 power supply is air-cooled so it becomes during operation.
When installing it, leave spaces of more than around it. Also make sure
to distribute cables away from the power supply.
6. Be sure to set the main switch to "
before replacing the light source bulb. (P. 26)
7. Always use the power cord provided by Olympus. If no power cord is
provided, please select the proper power cord by referring to the chapter
"PROPER SELECTION OF THE POWER SUPPLY CORD" at the end of
this instruction manual. If the proper power cord is not used, product
safety and performance cannot be guaranteed.
8. Always ensure that the grounding terminal of the microscope and that
of the wall outlet are properly connected. If the equipment is not grounded,
Olympus can no longer warrant the electrical safety and performance of
the equipment.
" (OFF) and unplug the power cord