Olympus IX81 Brochura e especificações - Página 15
Procurar online ou descarregar pdf Brochura e especificações para Microscópio Olympus IX81. Olympus IX81 16 páginas. Ix2 series research inverted system microscope
IX71 specifications
Microscope body
Revolving nosepiece
Focus
Primary image port
Frontal operation
Transmitted light
100 W transmitted light
IX2-ILL100
illuminator
illumination pillar
External power supply unit
TH4-100/200
Observation tube
Tilting binocular tube
U-TBI90
Binocular tube
U-BI90CT
U-BI90
Trinocular tube
U-TR30H-2+IX-ATU
Stage
Cross stage with flexible right handle
IX2-SFR
Cross stage with short left handle
IX-SVL2
Plain stage
IX2-SP
IX-MVR
Narrow plain stage
IX2-KSP
CK40-MVR
Gliding stage
IX2-GS
Condenser
Long working distance universal
IX2-LWUCD
Long working distance Relief Contrast
IX2-MLWCD
Ultra long working distance
IX-ULWCD
Water immersion DIC
IX2-DICD + IX2-TLW Single position for optical device (include two optical device holders),
Eyepiece
WHN10X
WHN10X-H
Reflected light
Fluorescence illuminator
IX2-RFAL
fluorescence unit
IX2-RFA
Fluorescence cube turret
IX2-RFAC
Light source
IX81 specifications
Microscope body
Revolving nosepiece
Focus
Primary image port
Frontal operation
Transmitted light
100 W transmitted light
IX2-ILL100
illuminator
illumination pillar
External power supply unit
IX2-UCB2
Observation tube
Tilting binocular tube
U-TBI90
Binocular tube
U-BI90CT
U-BI90
Trinocular tube
U-TR30H-2+IX-ATU
Stage
Cross stage with flexible right handle
IX2-SFR
Cross stage with short left handle
IX-SVL2
Plain stage
IX2-SP
Narrow plain stage
IX2-KSP
CK40-MVR
Gliding stage
IX2-GS
Condenser
Motorized long working
IX2-LWUCDA2
distance universal
Long working distance Relief Contrast
IX2-MLWCD
Eyepiece
WHN10X
WHN10X-H
Reflected light
Fluorescence illuminator
IX2-RFAL
fluorescence unit
IX2-RFA
Fluorescence cube turret
IX2-RFACA
Light source
IX81-ZDC specifications
Focusing position
Dry objective
Interface between air and cover glass
Oil immersion objective
Interface between sample (cultured liquid) and cover glass
Offset method
Controlled by software
Compensation for shift of observation position toward the focusing plane is by Z-axis control (built into the IX81-ZDC)
Observation methods
Fluorescence /DIC: DIC cannot be used beside gray-sensitive colors.
Dichromatic mirror IN/OUT method for AF laser introduction
Manual exchange
F.N. limitation
Light volume is low at the image perimeter for F.N. 22 when using 2X, 4X, 10X objectives
Focusing speed
Within approx. 0.8 seconds (average) from near focusing position)
(not including offset time through software
Focusing accuracy
±0.3 µm (when environmental temperature change is within 5°C)
Laser safety standard
Class 1 (JISC6802, IEC825, CDRH)
Laser safety function
Front monitor method (Laser light volume by special PD)
Camera port
Left side port
Can only be combined with U-TV1X-2+U-CMAD3, U-DPCAD, U-SIP (primary image , 1X)
Observation tube
IX-ATU+U-TR30H-2+IX-TVAD+U-CMT
IX-ATU+U-TR30-2+U-TV1X-2+U-CMAD3
27
Sextuple, simple waterproof mechanism incorporated
9 mm stroke (from stage surface, 7 mm upward and 2 mm downward), coaxial coarse and fine focusing knobs (minimum fine
focus graduation: 1 µm, full rotation of fine focusing knobs: 100 µm), upper limit stopper, torque adjustment for coarse focusing
Lower port (standard left side port: S1F 100% or S8F 80%, or optional lower Back port selectable, 2-step light path selection),
Upper port when built-in magnification changer 1X/1.6X is replaced (optional right side port),
Bottom port (option)
Light path selector, Transmitted light intensity control and light ON/OFF switch, TTL Pulse control switch
Pillar tilt mechanism (30° inclination angle, with vibration reducing mechanism),
Condenser holder (with 50 mm stroke, swing-in/out mechanism),
Field iris diaphragm adjustable, 4 filter holders (ø45 mm, t=6 mm or less)
Two versions available (100 V and 200 V), Optional TH4-HS hand switch can be used, 2.2 kg weight
35-85° continuous angle adjustable (eyepoint height range: 406 mm-471 mm),
interpupillary distance adjustable between 50-76 mm, diopter adjustment function, erect image, F.N. 22
Built-in focusing telescope, interpupillary distance adjustable 50-76 mm, diopter adjustment function, F.N. 22
Interpupillary distance adjustable 50-76 mm, diopter adjustment function, F.N. 22
3 step optical path selectable (observation : straight port = 100:0, 20:80, 0:100),
interpupillary distance adjustable 50-76 mm, diopter adjustment function, F.N. 22
50 mm (X) X 50 mm (Y) stroke, stage insert plate exchangeable (ø110 mm)
50 mm (X) X 43 mm (Y) stroke, stage insert plate exchangeable (ø110 mm)
232 mm (X) X 240 mm (Y) stage size, stage insert plate exchangeable (ø110 mm)
Mechanical stage to be used with IX2-SP, 130 mm (X) X 85 mm (Y) stroke
160 mm (X) X 240 mm (Y) stage size, stage insert plate exchangeable (ø110 mm)
Mechanical stage to be used with IX2-KSP, 120 mm (X) X 78 mm (Y) stroke
Upper circular stage 360° rotatable, 20 mm (X/Y) travel
5 positions for optical devices (3 positions for ø30 mm and 2 position for ø38 mm),
aperture iris diaphragm adjustable, N.A. 0.55 / W.D. 27 mm
4 positions for optical devices (for ø50 mm, Relief Contrast optical devices rotatable),
aperture iris diaphragm adjustable, N.A. 0.5 / W.D. 45 mm
4 positions for optical devices (for ø29 mm), aperture iris diaphragm adjustable, N.A. 0.3 / W.D. 73 mm
40° injection pipette or electrode insertion angle, aperture iris diaphragm adjustable, N.A. 0.9 / W.D. 3.7 mm
High eyepoint, F.N. 22
High eyepoint, diopter adjustment function, F.N. 22
L-shaped design with exchangeable F.S. and A.S. modules, two filter holder sliders (2 positions, ø32 mm, t=6 mm or less)
Straight design with field iris diaphragm, filter holder slider (2 positions, ø32 mm, t=6 mm or less)
6 positions in a rotating turret, built-in shutter
100 W Hg lamp housing and transformer, or 75 W Xe lamp housing and transformer
Sextuple motorized with objective lens retraction in PC mode,
simple waterproof mechanism incorporated
9 mm stroke (from stage surface, 7 mm upward and 2 mm downward), fine/coarse switchable focusing knobs (minimum
graduation: 0.01 µm), objective lens escape/return buttons and return to memory position buttons (each side of microscope
frame)
Lower port (standard left side port: S1F 100% or S8F 80%, or optional lower Back port selectable, 2-step light path selection),
Upper port when built-in magnification changer 1X/1.6X is replaced (optional right side port),
Bottom port (option)
Light path selector button, Transmitted light intensity control buttons and light ON/OFF switch button, Fine/Coarse focus
selector button, TTL Pulse control switch (auxiliary) buttons
Pillar tilt mechanism (30° inclination angle, with vibration reducing mechanism),
Condenser holder (with 50 mm stroke, swing-in/out mechanism), Field iris diaphragm adjustable,
4 filter holders (ø45 mm, t=6 mm or less)
Auto voltage selector (100 V / 200 V), RS232C interface for PC operation, IX2-BSW driver software
35-85° continuous angle adjustable (eyepoint height range: 406 mm-471 mm),
interpupillary distance adjustable between 50-76 mm, diopter adjustment function, erect image, F.N. 22
Built-in focusing telescope, interpupillary distance adjustable 50-76 mm, diopter adjustment function, F.N. 22
Interpupillary distance adjustable 50-76 mm, diopter adjustment function, F.N. 22
3 step optical path selectable (observation: straight port = 100:0, 20:80, 0:100), interpupillary distance adjustable 50-76 mm,
diopter adjustment function, F.N. 22
50 mm(X) X 50 mm(Y) stroke, stage insert plate exchangeable (ø110 mm)
50 mm(X) X 43 mm(Y) stroke, stage insert plate exchangeable (ø110 mm)
232 mm(X) X 240 mm(Y) stage size, stage insert plate exchangeable (ø110 mm)
160 mm(X) X 240 mm(Y) stage size, stage insert plate exchangeable (ø110 mm)
Mechanical stage to be used with IX2-KSP, 120 mm (X) X 78 mm (Y) stroke
Upper circular stage 360° rotatable, 20 mm(X/Y) travel
Motorized turret with 6 position slots for optical devices (3 positions each for ø30 mm and ø38 mm),
aperture iris diaphragm adjustable, N.A. 0.55 / W.D. 27 mm
4 positions for optical devices (for ø50 mm, Relief Contrast optical devices rotatable), aperture iris diaphragm adjustable,
N.A. 0.5 / W.D. 45 mm
High eyepoint, F.N. 22
High eyepoint, diopter adjustment function, F.N. 22
L-shaped design with exchangeable F.S. and A.S. modules, two filter holder sliders (2 positions, ø32 mm, t=6 mm or less)
Straight design with field iris diaphragm, filter holder slider (2 positions, ø32 mm, t=6 mm or less)
Motorized turret with 6 positions, built-in shutter
100 W Hg apo lamp housing and transformer, 100 W Hg lamp housing and transformer or
75 W Xe lamp housing and transformer
Speed also varies according to the start position of auto focusing, and
individual PC performance
IEC60825
IX71 dimensions
*243~391
135
394
290
Weight: 16 kg Power consumption: 200 VA
* Lengths with an asterisk (*) vary according to interpupilllary distance.
IX71+fluorescence illuminator combination dimensions
*243~391
135
394
290
584
Weight: 20.2 kg Power consumption: 480 VA
* Lengths with an asterisk (*) vary according to interpupilllary distance.
IX81-ZDC dimensions
368
4
5
3
135
394
290
584
(unit: mm)
IX81 dimensions
*243
~
391
135
394
290
Weight: 16.9 kg Power consumption: 350 VA
* Lengths with an asterisk (*) vary according to interpupilllary distance.
(unit: mm)
IX71+L-shaped fluorescence illuminator dimensions
*243~391
135
290
549
Weight: 21.2 kg Power consumption: 480 VA
* Lengths with an asterisk (*) vary according to interpupilllary distance.
(unit: mm)
TH4 dimensions
75
200
14.5
IX2-UCB2 dimensions
RMT
ERR
IX-UBC
NP
MU
RSHT
AS
FW1
FW2
FW3
TL
CDT
Z/AF
SW1
ON
OFF
SW2
RS232C
See manual
HS
15
125
310
332 (depth)
U-HSTR2 dimensions
OB
180
7°
U-FH dimensions
54.0
28.3
35.5
82.3
70.0
4.0
75.5
(unit: mm)
(unit: mm)
394
453
(unit: mm)
28