Olympus MX61A Manual - Página 7
Procurar online ou descarregar pdf Manual para Equipamento de laboratório Olympus MX61A. Olympus MX61A 48 páginas. Motorized 300 mm compatible semiconductor/fpd inspection microscope
MX61A external view, eye point and center of gravity: When the 8x8-inch stage is combined
Eye point
(Bottom view of microscope)
(Note) The center of gravity is an approximate position when the microscope is equipped with the standard module
combination for transmitted light observation. Note that the position is variable depending on the weight of speci-
men, position of the stage and other modules used.
9. To prevent toppling of the microscope system, keep the total height of the microscope below 1 meter (3.3 ft) when
attachments (including Olympus optional modules and the CCD camera prepared by the customer) are mounted.
@
Fig. 1
10. Each microscope has screw holes (M5, depth 10 mm) for prevention of
toppling (in the case of an earthquake or microscope imbalance) on the
²
side panels @ (x 2 holes) and the rear panel ² (x 2 holes). Clamp the
microscope using L-shaped clamps and these screw holes as required.
When clamping the microscope using L-shaped clamps prepared
by the customer, be sure to use steel bolts (strength category 12.9)
with as long as possible threaded sections (8 mm or more is rec-
ommended).
Unit: mm
Position of center of gravity
MX61A
4