Olympus MX61A Manual - Página 9

Procurar online ou descarregar pdf Manual para Equipamento de laboratório Olympus MX61A. Olympus MX61A 48 páginas. Motorized 300 mm compatible semiconductor/fpd inspection microscope

1
Getting Ready
@
²
Fig. 2
2
Maintenance and Storage
1. To clean the lenses and other glass components, simply blow dirty away using a commercially available blower and wipe
gently using a piece of cleaning paper (or clean gauze).
If a lens is stained with fingerprints or oil smudges, wipe it gauze slightly moistened with commercially available absolute
alcohol.
Since the absolute alcohol is highly flammable, it must be handled carefully.
Be sure to keep it away from open flames or potential sources of electrical sparks –– for example, electrical
equipment that is being switched on or off.
Also remember to always use it only in a well-ventilated room.
2. If any part of the equipment (other than glass components) gets dirty, with it with a clean cloth.
If the part is extremely dirty, do not attempt to use organic solvents to clean it; instead, use a soft, lint-free cloth slightly
moistened with a diluted neutral detergent.
3. Never disassemble any part other than instructed of the microscope. This could result in malfunctions or reduced
performance.
4. When not using the microscope, keep it covered with a dust cover. Make sure the lamp housing is cool before covering
the microscope.
5. When disposing of the microscope. Check the regulations and rules of your local government and be sure to observe
them.
1. Do not use the microscope where it is subjected to direct sunlight, high
temperature and humidity, dust or vibrations. (For the operating conditions,
refer to the INSTRUCTIONS manual's chapter 8, "SPECIFICATIONS" on
page 43.)
2. A microscope is a precision instrument. Handle it with care and avoid
subjecting it to sudden or severe impacts. Never attempt to rotate the
motorized revolving nosepiece by hand; otherwise, the gear or other parts
may be damaged.
3. When moving the microscope, detach the observation tube, stage, lamp
housing and specimen to reduce the total system weight, then insert the
two provided carrying rods @ firmly into each of the left and right side
panels, and attach the two provided guarding parts ² on the two carrying
rods on the front.
If the guarding parts are not attached, the person holding the front
carrying rods may have a finger caught between the carrying rod
and the desktop.
Two persons are needed to carry the microscope; one should hold the
two front carrying rods and the other person should hold the rear carrying
@
rods. Be sure to move the microscope cautiously. (The microscope frame
weight is about 31 kg, or 68.2 lbs.)
}After movement, remove the guarding parts and carrying rods and retain
them carefully without losing. The carrying rods can be inserted in the
four threaded holes on the rear of the microscope. For the screw holes
left by removing the carrying rods, attach the provided screw caps on the
holes for the rear carrying rods and the cap seals on the holes for the
front carrying rods.
# Do not change the position of the microscope by sliding it on the
desktop surface; otherwise, the rubber feet will be damaged.
4. Be sure to attach the transport clamping plate(s) to the stage before
transporting it.
With the 14x12-inch stage combination, the center of gravity varies
greatly depending on the position of the stage. To ensure extra
stability, keep the two provided carrying rods attached to the mounting
holes on the front right side when using the microscope.
MX61A
6