First, remove
the stage mounted
on the BX-series
microscope,
replacing
it with the U-SRG rotatable
stage. When replacing
the stage, be very careful to avoid striking
parts such as objectives.
Fig. 1
Circle described
by specimen
mark
/ Center
Fig. 2
Fig. 3
0
To
make
centering
easier, use at eyepiece
with crosshairs.
1. Move the 1 OX objective
into the light path and place an object that
can serve as a specimen
mark at the center of the crosshairs.
2. Rotate the stage and use the two stage centering
knobs 0 to move
the center of the circle described by the specimen
mark to the center
of the crosshairs.
(Fig. I)
* Repeat the above
procedure
several
times
until the stage is
accurately
centered.
(Fig. 2)
3. Insert the specimen
clamped @ into whichever
holes in the stage are
most convenient.
(Fig. 3)