Fig. 11
33
²
When the mounting position is 5, 6 or 7:
}To clamp the DIC prism, do not use the built-in optical ele-
ment centering screw but use the provided Allen wrench ²
which is designed exclusively for this purpose.
1. Engage the position to be used in the light path by motorized
operation, loosen the position clamping screw by inserting the
Allen wrench ² from the rear of the condenser, then remove
the dummy optical element.
# Do not turn the turret while the clamping screw is loose.
Otherwise, the clamping screw may interfere with other
part.
2. Drop in the DIC prism by aligning its positioning pin with the
slot, then tighten the screw using the Allen wrench. Then
attach the index plate.
When all of the required optical elements have been mounted,
place the turret cover in the original position.