Inspection and maintenance
2.2.4

Tilt drive

SIM
Actuator cams
SIE
Safety limit switch
Push the actuator cams for safety limit switches s31 and s32 by hand. When the switch
is activated, the K1 contactor in M20 drops.
Check that the actuator cam moves easily.
SIE
Travel range
Move the system over the entire travel range of +88° to -15°and back checking for noise
and ease of movement.
2.2.5

Rotating column drive

SIE
Safety limit switch
SIM
Actuator disk
Position the table horizontally.
Turn the actuator disk for the safety switch S41 right and left by hand until the switch
activates (the contactor K1 in M20 must drop. Also check that the actuator disk moves easily.
SIE
Travel range
Move the system over the entire travel range from +15° to -15° and back checking for
noise and ease of movement.
2.2.6

Longitudinal table drive

SIM
Actuator cams
SIE
Safety limit switch
Activate safety limit switches s71and s72 by hand.When the switch is activated, the K1
contactor in M20 drops.
In both end positions, the actuator cams must stop approximately 5 mm before the
respective switch.
SIE
Travel range
Measure the travel range of 573 cm toward the physician's side and 227 cm in the
headward direction (from the edge of the tabletop on the physician's side to the white
plastic piece in the table rail) and check for noise and smooth travel.
Siemens AG
Medical Engineering
RLL5-310.101.02
Page 3 of 16
Rev. 02
01.97
TDSP 1
2 - 3
UROSKOP D