Inspection and maintenance
2.2.9

System general

SIE
Main operating console
Check all functions on the main operating console.
SIE
Footswitch
Check the exposure and fluoro functions of the footswitch with the SS switch off.
SIE
IR footswitch
Check all functions of the IR footswitch.
SIE
Tableside control
Check all functions of the tableside control unit making sure to connect the unit to the X-
ray tube arm as well as to the spotfilm device.
SIM
Accessories
Check all accessories to be sure that they mount securely on the system.
SIM
Cables and corrugated tubing
Check that all the cables and corrugated tubing are in good condition, if they are
accessible.
SIM
Cables and corrugated tubing
SIM
Strain relief
Where accessible, check all strain reliefs on the cables and corrugated tubing for
correct position and secure mounting.
SIM
Cable mounts
Check all cable mounts where accessible for the cables and corrugated tubing for
correct position and secure mounting.
2.2.10

Safety devices

SIE
Emergency OFF switches
Press the emergency OFF switches in sequence and individually:
- at the spotfilm device
- on the main control console
- on the X-ray tube arm
When the emergency STOP switch is activated:
- Error U01 must appear in the area dose product display.
- Error 0050 must appear on board D1 in M20.
- system movements must be blocked.
Unlock the emergency OFF switches on the system
- the system must automatically reset.
Siemens AG
Medical Engineering
RLL5-310.101.02
Page 5 of 16
Rev. 02
01.97
TDSP 1
2 - 5
UROSKOP D