Siemens UROSKOP D Instruções adicionais de instalação, operação e manutenção - Página 13
Procurar online ou descarregar pdf Instruções adicionais de instalação, operação e manutenção para Equipamento industrial Siemens UROSKOP D. Siemens UROSKOP D 26 páginas.
Inspection and maintenance
2.2.9
System general
SIE
Main operating console
•
Check all functions on the main operating console.
SIE
Footswitch
•
Check the exposure and fluoro functions of the footswitch with the SS switch off.
SIE
IR footswitch
•
Check all functions of the IR footswitch.
SIE
Tableside control
•
Check all functions of the tableside control unit making sure to connect the unit to the X-
ray tube arm as well as to the spotfilm device.
SIM
Accessories
•
Check all accessories to be sure that they mount securely on the system.
SIM
Cables and corrugated tubing
•
Check that all the cables and corrugated tubing are in good condition, if they are
accessible.
SIM
Cables and corrugated tubing
SIM
Strain relief
•
Where accessible, check all strain reliefs on the cables and corrugated tubing for
correct position and secure mounting.
SIM
Cable mounts
•
Check all cable mounts where accessible for the cables and corrugated tubing for
correct position and secure mounting.
2.2.10
Safety devices
SIE
Emergency OFF switches
•
Press the emergency OFF switches in sequence and individually:
- at the spotfilm device
- on the main control console
- on the X-ray tube arm
When the emergency STOP switch is activated:
- Error U01 must appear in the area dose product display.
- Error 0050 must appear on board D1 in M20.
- system movements must be blocked.
•
Unlock the emergency OFF switches on the system
- the system must automatically reset.
Siemens AG
Medical Engineering
RLL5-310.101.02
Page 5 of 16
Rev. 02
01.97
TDSP 1
2 - 5
UROSKOP D