Programming the Thawing Time. When the thawing time is displayed, press MINUS because you want to
change the time. Use PLUS or MINUS to set the desired new thawing time, press ENTER to program in the
new time. Turn the unit off.
Calibrating Temperature. Turn the unit on and let it heat up to 37.0
your accurate thermometer. There is a covered THERMOMETER WELL on the left side of the control panel.
Rotate the cover bar to expose the thermometer well. Place your thermometer into the THERMOMETER
WELL. The thermometer stem should be at least 15 cm. (6") long. If the displayed temperature needs to be
changed, turn the unit off. Enter PROGRAMMING MODE by holding ENTER as you turn the unit on. When
the thawing time is displayed push PLUS until the display shows "CALIBRATE TEMPERATURE". Push
MINUS because you want to change the displayed temperature. Use PLUS or MINUS to make the displayed
temperature agree with your thermometer. Push ENTER to record the calibration. Turn the unit off. If the
unit can not be calibrated, or if the calibrated temperature is not stable, have a technician check the unit. See
TECHNICIAN FUNCTIONS.
Adjust the massaging. Pump 1 fills the outer tubes of the BLADDER. Pump 2 fills the center tube BLADDER.
Pump 1 is on constantly when thawing, but at a lower pressure. Pump 2 then controls the massaging action:.
1. Pump 2 ON time pressurizes the center tube and pushes the plasma towards the outside tubes.
2. Pump 2 OFF time lowers the pressure of the center tube and allows plasma to be pushed to the center.
You can observe the action and adjust the Pump 2 times to change the massaging action. The preset times
have been tested to work very well.
SWITCH FUNCTIONS
There are 6 switches on the front panel:
START. Starts function.
STOP. Stops function.
ENTER. Records parameter changes.
PLUS. Steps through parameters and increases value of parameter.
MINUS. Allows change in parameter and decreases value of parameter.
Sketch of FIN logo. Not currently used. Future expansion
o
C. Check the displayed temperature with