7. Calibration
Before calibration, You should select Dip switch to the most near thing in sensor
output value and make
a)Sensor Supply Voltage (ON : 5V, OFF : 10V)
b)Amplification Adjustment Switch(1.0mV/V, 1.5mV/V, 2.0mV/V, 3.0mV/V)
(Set with the number which gets near to maximum output value of sensor .)
Press Function key about 3 seconds to return to Setup mode.
Press Up key to be
Press ENT key to choose Sensor Supply Voltage mode.
Press Up key to choose what you want.
Press ENT key to store it.
->
If you want to get out of the setting mode and return to the measurement mode,
you may have to push the Function key.
7-1. Real-weight Calibration
Standard weight is necessary. When being used Multiple sensor at one Indicator, It
calibtrate certainly at Real-weight.
1) Real-weight Calibration Mode Setting
Press Function key about 3 seconds to return to Setup mode.
Press Up key to be
->
Press ENT key to be
※ If, when being in
be
.
Press ENT key to store it.
->
If you want to get out of the setting mode and return to the measurement mode,
you may have to push the Function key.
.
->
->
(In case it is not good,
11
, press
Up key to
will come out.)