7. Calibration
Before calibration, You should select Dip switch to the most near thing in sensor
output value and make
a)Sensor Supply Voltage (ON : 5V, OFF : 10V)
b)Amplification Adjustment Switch(1.0mV/V, 1.5mV/V, 2.0mV/V, 3.0mV/V)
(Set with the number which gets near to maximum output value of sensor .)
Press Function key about 3 seconds to return to Setup mode.
Press Up key to be
→
Press ENT key to choose Sensor Supply Voltage mode.
Press Up key to choose what you want.
Press ENT key to store it.
->
If you want to get out of the setting mode and return to the measurement mode,
you may have to push the Function key.
7-1. Real-weight Calibration
Standard weight is necessary. When being used Multiple sensor at one Indicator, It
calibtrate certainly at Real-weight.
1) Real-weight Calibration Mode Setting
Press Function key about 3 seconds to return to Setup mode.
Press Up key to be
->
Press ENT key to be
※ If, when being in
be
.
Press ENT key to store it.
->
If you want to get out of the setting mode and return to the measurement mode,
you may have to push the Function key.
.
->
->
(In case it is not good,
11
, press
Up key to
will come out.)