dacell DN-15W Руководство пользователя - Страница 14

Просмотреть онлайн или скачать pdf Руководство пользователя для Измерительные приборы dacell DN-15W. dacell DN-15W 16 страниц. Digital indicator

Press ENT key
Press ENT key to set Shunt CAL data.
->
If you want to get out of the setting mode and return to the measurement mode,
you may have to push the Function key.
※ when turn on Shunt CAL S/W, displays Shunt CAL data
and when turn off Shunt CAL S/W, displays " 0 ".
If mistakes did occur , repeat the above statement.
※ note) Please turn off SHOUT CAL S/W after calibrating.
7-2. Constant value Calibration
Sensor CAPA and Output value are necessary. (When the sensor purchasing, it is
becoming marking in Calibration sheet.)
It considers a sensor quality and a circumference condition, use Real-weight
calibration for precision measuring.
1) Constant value Calibration mode setting
Press Function key about 3 seconds to return to setup mode
Press Up key to be
->
Press ENT key to set sensor output value.
After selecting from in
Press ENT key to store it.
->
If you want to get out of the setting mode and return to the measurement mode,
you may have to push the Function key.
(Currently the value is indicated.)
※ Set with the number which gets near to maximum output value of sensor.
2) CAPA setting and Sensor output setting
Set it with reference to Calibration sheet, enclosed with the sensor.
Press Up key about 3 seconds to return to CAPA mode.
Press ENT key, It uses
value.
, this then turn on Shunt CAL S/W in back panel
->
->
->
(In case it is not good,
Shift key and
13
will come out.)
Up Key and it inputs Sensor CAPA