6. Installation and mounting

6.1 Training

Installation should only be conducted by qualified service person-
nel according to local regulations or Temposonics trained service
technicians. Temposonics offers web based and in person training for
installation, commissioning, maintenance, and repair. Temposonics
also offers factory direct services for these same functions. Contact
Temposonics to discuss training or factory direct services before
starting.

6.2 Stilling wells and guide poles

Level Plus
transmitters can be mounted in slotted or unslotted stilling
®
wells but a slotted stilling well is always preferred. Using a unslotted
stilling well will negatively affect performance of any level device as
the level in the stilling well can differ from the level in the tank. The
Level Plus
transmitter can also be installed to one side of the stilling
®
well to also allow for sampling and manual gauging from the same
opening as the automatic tank gauging. Contact technical support for
details.
Level Plus
transmitters do not require a stilling well for installation.
®
Our transmitters are installed in numerous tanks without stilling wells
with no loss in performance due to our patented flexible waveguide
and hose. A stilling well is highly recommended for agitated, turbulent,
and/or fast filling tanks.

6.3 Tools

• Adjustable pliers
• Large pipe wrench
• Common head screwdriver, slotted screwdriver

6.4 Installation steps

Caution:
It is recommended that assembly and mounting of this transmitter
should not be done alone. To ensure proper and safe assembly
of the level transmitter, a minimum of two (2) individuals are
recommended. Gloves are also recommended. PPE may be required
for work areas such as safety shoes, safety glasses, hard hat, and
fire resistant clothing.
1. Consult chapter 4.3 before starting.
2. Perform steps 1-10 in chapter 8.4.1
3. Remove stop collars from rigid pipe.
4. Slide NPT adapter or flange onto rigid pipe and secure the
adapter/flange. To prevent damage to rigid pipe keep the adapter/
flange supported and towards the end of the level transmitter.
5. Slide the product float onto the rigid pipe. Slide the interface float
(optional) onto the rigid pipe. Install stop collar so the magnet of
the float is 64 mm (2.5 in.) from the tip of the rigid. Do not drop
float(s) or allow them to free fall along the rigid pipe as damage
may result.
NOTICE
The bottom stop collar can be removed or adjusted based on the
float selected for the application. Please consult the factory for more
information.
6. Slide float(s) and adapter or flange assembly back down to the
stop collar to prevent them from free falling during installation
into the tank. If you are using the o-ring stop collar it will not
support the weight of the adapter or flange. Insert the rigid pipe
and floats through the tank opening and lower the transmitter
float assembly into the tank until it rests on the bottom.
7. Connect the adpter or flange to the mating connector on the tank.
8. Connect the 3/4" NPT (rigid) to the adapter or flange
assembly. Tighten the NPT fitting to the adapter or flange. It is
recommended to use anti-seize on the connection.
9. Connect cable to connector.
10. Terminate the pigtail cable noting proper wire orientation. Wiring
diagram is shown in Section 7.
Consult Section 5.7 Technical Drawing to reference what the finished
installation should look like.