Amprobe AMB-50 Руководство пользователя - Страница 15
Просмотреть онлайн или скачать pdf Руководство пользователя для Испытательное оборудование Amprobe AMB-50. Amprobe AMB-50 17 страниц. 5000v insulation resistance tester
SPECIFICATIONS
Safety and Compliances
Certification
�
Compliances
IEC 61010 CAT.III 600V overvoltage and double insulation standard
Physical Specifications
Display (LCD)
Operating Temperature
Storage Temperature
Relative Humidity
Battery Type
Dimensions (H x W x L)
Weight
General Specifications
Range
Overloading
Battery Indicator
Icon Display
Current Consumption
Feature Summary
Display Backlight
Computer connection
Data Logging and Recall
Autorange
Warning
Voltage
COMP Measurement
PI Measurement
TIME
Digital: 1999 counts
Analog bar graph.
0°C~40°C (32°F ~104°F )
-20°C ~60°C (-4 °F~152°F )
85% @ -10°C ~40°C below;
90% @ 0°C ~40°C :
8pcs X 1.5V (LR14) batteries or power adaptor (input voltage 230V, 50/60Hz,
75mA, input DC14V, 1.0A).
Power adaptor is not included
202 x 155 x 94 mm
Approx. 2kg (4.4 Lbs) (including battery)
Auto
Display OL on insulation resistance range
Display
Equips with function and battery indicator icons.
Maximum: around 1.0A
Average: around 20mA
Bright backlight for clear readings in poorly lighted areas.
Via USB interface.
18
The Meter automatically selects best range
and red light will on.
Auto release voltage
Use the Compare function to set a pass/fail compare level for the insulation
measurements.
Polarization Index is the ratio of insulation resistance. You can pre-set two point
of times and automatically carry out the measurement.
To carry out measurement by setting a specified time within 15 minutes.
10