6.2 How to Operate

6.2.1 Rotor Installation

The drive spindle in the chamber and the inside cones of rotor should be first
wiped with a piece of soft and clean cloth, then place the rotor with two hands
to the cone side of the drive spindle in the centrifuge chamber, Washer should
be placed before tightening the screw nuts.

6.2.2 Tubes Installation

Place the bracket into the slot of rotor; make sure that the bracket is turning
smoothly, tube sets (adaptors) are placed into holes of brackets, the solution to
be separated into the tubes. Visually check if the solution in the tubes is
roughly equal before putting them into the tube sets.

6.2.3 Preset

1) Function Declaration of Display Window
① Rotor number Display Window:
is on, the window of rotor number is flashing, turn the "ENTER" key to
confirm the number.
② Speed Dispaly Window:
Note: the set speed must not exceed the max. speed limit.
③Time Display Window:
reference to the following sheet 2.
This button can set rotor number, speed and time.
④ Set:
can set rotor number, speed and time.
2) Set button :
Fig.1 Control Panel
press "SET" ,when power indicator (red)light
It can display the running speed and RCF.
The window can display the time and the Erro, Detailed