Premium has two methods of controlling the fixture; the Keypad Controller (Shown in Figure 3) and the
®
Hand-held Remote Controller (Shown in Figure 4). These two controls mirror each other; treatment times entered into Keypad
Controller will appear on the Hand-held Remote Controller and vice versa.
u
K
sing tHe
eyPad
1.
Setting Treatment Time on the Keypad is done using the HOURS and MINUTES buttons which is located on the top of the
Sanidyne
Premium. (See Figure 1 for location) Selecting the up or down facing arrows located on the keypad in line with
®
the HOURS or MINUTES button will either increase or decrease the hours or minutes. The hours can be set between 0 and
24, and the minutes can be set from 1 to 59. (Example of one hour and five minutes would appear as 01:05 on the Treatment
Time Display). Since the Keypad Controller will no longer be viewable once the treatment time begins, it is recommended that
the Hand-held Remote Controller be fully charged and powered before this time and be in possession of the operator
at all times.
2. Determining the proper treatment times for your specific location can be done by using the "Recommended Treatment Time
for Room Size" chart (Figure 5 on Page 13).
3. Once the choice of time has been displayed on the Keypad controller, select the START/STOP button to begin and leave the
Operation Area immediately. Make sure that the Hand-held Remote Controller is kept in possession of the operator during the
Treatment Time.
Once you select the START/STOP button an audible exit alarm will begin to sound and the Sanidyne
will begin counting down from 1:30 seconds to allow all inhabitants to leave the area before the UV treatment begins.
Additionally, the word "PrEP" will begin flashing on the Keypad and the Occupancy Sensors located on the top of the unit
(Shown on Figure 1) will begin to operate at this time, providing an additional safety area around the unit. As an additional
safety feature, if the sensors detect movement close to the Sanidyne
countdown time, the Sanidyne
Occupancy Sensors do not protect the entire Operation Area of the Sanidyne
circumstances should this sanitizer be permitted to operate with humans, plants, or animals present in the Operation Area.
4. Outside the Operation Area the Treatment Time can be viewed on the Hand-held Remote Controller. Once the treatment time
has expired the Sanidyne
return to the area. The word "donE" will begin flashing on the Keypad Controller on the top of the fixture.
u
H
-
sing tHe
and
Held
The Hand-held Remote Controller provides further detailed information on the Sanidyne
recommended to review the Hand-held Remote Controller Overview section of this manual to become familiar with the display
which is shown in Figure 4.
The Hand-held Remote Controller can operate the Sanidyne
imperative that the operator make a visual inspection of the Operation Area to ensure it is clear of all humans, plants, or
animals before proceeding.
1. Setting Treatment Time on the Hand-held Remote Controller is done by holding a finger on the display face, directly on the
words Treatment Time until the screen changes to a scroll bar. Next scroll the numbers up or down, characters on the left are
hours, characters on the right are minutes. Positioning the desired treatment time in the center row. (Example of one hour and
five minutes would appear as 01:05 on the Treatment Time Display) When times are positioned correctly, select DONE on the
bottom of the scroll bar to input your Treatment Time choice.
2.
Determining the proper treatment times for your specific location can be done by using the "Recommended Treatment Time
for Room Size" chart (Figure 5 on Page 13).
3.
Once the choice of time has been displayed on the Hand-held Remote Controller, select the start button to begin and leave the
Operation Area immediately. Make sure that the Hand-held Remote Controller is kept in possession of the operator during the
Treatment Time. (Remote Controller Instructions Continued on Page 12)
o
Peration
C
s
ontroller to
et
Premium will automatically reset giving 30 seconds more to leave the Operation Area.
®
Premium will turn off the STER-L-RAY
®
r
C
emote
ontroller to
t
t
reatment
imes
Premium when less than 30 seconds remain on the
®
Germicidal Ultraviolet Lamps and it is now safe to
®
s
t
et
reatment
Premium
®
Premium. Under no
®
t
imes
Premium. Therefore it is
®
Premium remotely, thus it is
®