DataApex Shodex RI-101 Руководство пользователя - Страница 11
Просмотреть онлайн или скачать pdf Руководство пользователя для Лабораторное оборудование DataApex Shodex RI-101. DataApex Shodex RI-101 18 страниц. Clarity controls
4 Using the control module
4.1 Method Setup - Acquisition
The
Method Setup - Acquisition
parameters of the Shodex RI-101 detector. If more than one detector is
available, it is possible to select between them by using the Select
Detector combobox on the top of the dialog.
Sampling Period [s]
Defines the frequency of data being gathered in the detector. The
frequency of the data being sent to Clarity from the detector is defined by
Response Time parameter. Valid values for the Shodex RI-101 detector
range from 0.1 to 2.0 s.
Range for Integrator Output [µRIU/V]
Sets the output range for the Integrator Output analog output, as well as
digital signal sent to Clarity . The available range is dependent on the
Detector Model as set in the
Range for Recorder Output [µRIU/FS]
Sets the output range for the Recorder Output analog output, as well as
digital signal sent to Clarity.
Polarity of the Output Signals
Determines the polarity of the output signal.
Balance of the Output Signals [mV]
Defines the value to which the signal level will be reset when Autozero
function is used.
Fig 5: Method Setup - Acquisition
DataApex UNI Setup
tab serves for setting the common
dialog.
- 7 -
Clarity Controls