ADE M400020 Series Руководство по эксплуатации - Страница 9
Просмотреть онлайн или скачать pdf Руководство по эксплуатации для Весы ADE M400020 Series. ADE M400020 Series 16 страниц. Class iii approved scales electronic chair scales
Weighing
Press the ON key for at least 3 seconds to turn the scales on. An automatic function test is carried out after
being switched on. This test is completed when the display shows zero. The scale is now ready for the weigh-
ing procedure.
Resetting the scales back to zero
Wait until the standstill status indicator
(14) appears. By pressing the ZERO key (8), measuring inaccura-
cies after switching on the scale are being compensated. The zero key restarts the scale. The status indicator
(6) appears when the scale is being resetted.
Tare function
The TARE key (9) is used when the patient is being weighed with other items (e.g. clothing, shoes or other
items) which should be ignored by the scale. To do this, weigh these items first. After the weight is displayed,
press the TARE key (9) to reset the display back to zero. The display changes to net mode (NET status indica-
tor (5) is on). The weight of the previously weighed and saved items is ignored when the patients' weight is
displayed. Enter the tare button again to get back to the normal weighing mode and remove any load from the
platform.
Entering manually tare - values:
Enter the appropriate tare value just by using the key pad under normal weighing mode. If the tare value has
been entered correctly, the value will be stored. Press the ENTER key (or TARE). The ADE-display is now
ready for weighing and changes to net mode [NET status indicator (5) is on]. Enter the tare button again to
leave this mode.
The manual tare value must be >0, the maximum value is limited by the weighing range.
Preset tara
Press and hold the PRESET TARE key (10) [approx. 3 sec.], PRO PT X appears on the display. Select the
memory location 1-10. Now enter the TARE VALUE which you would like to be ignored by the scale and con-
firm same with ENTER. Enter the tare button again to leave this mode.
Restore the preset tara value:
Press the PRESET TARE key (10). "Pt x" appears on the display. Enter a number from 0-9 (memory location
for the tare value). If no number is entered within five seconds, the unit will turn back to weighing mode.
After entering the number, the ADE-MAK accesses the preset tare value, shows it for five seconds on the dis-
play and then changes to net mode. The accessed tare value is accepted at the same time. Enter the tare
button again to leave this mode.
Displaying the gross weight, net weight and tare value:
Press the "0" key to change between displays for tare, gross weight (NET status indicator (5) off) and net
weight (NET status indicator (5) on). If a tare value is saved, the net weight shown corresponds to the gross
weight minus the tare value. If there is no tare value saved, the display remains in gross mode.
The tare value flashes briefly, then after a second the display changes to gross mode.
Hold function
Use the HOLD key to hold the display if there are weight fluctuations which e.g. are caused by patient move-
ments on the scales. You could also use this function if you like to hold the results at the display after the
weight/load has been removed from the scale.
To do this, the scales must be at standstill, e.g. the standstill status indicator
(14) is on (depending on the
setting it is not always necessary for the scales to be at standstill). The Hold status indicator (15) is on.
Deactivate the HOLD function by pressing the key again.
M400020-210322-Rev007-UM-e
9