Information
?
Intended Purpose
To support the patient's arms overhead with
controlled shoulder flexion for imaging.
Intended for bore's 70cm and upward.
Intended User/Training Requirement
Intended to be used by trained medical
professionals.
Patient Target Group
Adult male or female patient requiring
diagnostic imaging and image guided therapy.
!
Contraindications
Not to be used with patients over 135Kg.
Not for paediatric use.
Not intended for bore size less than 70cm.
!
Warnings and Cautions
Ensure IFU is read prior to use.
Ensure Preparatory Cleaning is
conducted.
Ensure Preparatory Product Check is
conducted.
Ensure MRI safety checks are conducted.
Ensure patient and device will fit clear of
the bore.
Do not use with patients over 135Kg.
Not for paediatric use.
Ensure general patient insulation protocol
is followed for MRI.
For CT and MRI, ensure there is adequate
patient clearance prior to use. (refer to
page 8)
Incident Reporting Guidelines
For product complaints and incidents, please
complete form on
adeptmedical.co.nz
/repairs
12
adeptmedical.com |
|
Initial Checks & Cleaning
Cleaning
Refer to Disinfection Instructions.
Product Checks
No visible damage or sharp edges, e.g.
cracks.
Arm Strap: not damaged, e.g. tearing.
Locking Leg Assembly: ensure that
the Locking Leg Assembly is not free to
move in the locked position.
Locking Leg Assembly: Ensure it can
move up and down when pushed in the
unlocked position.
Ensure the Locking Leg Assembly is
fully inserted into the Daggerboard and
the release clip is fully engaged (refer to
page 5).
Ensure the Armrest Wing is fully inserted
onto the Locking Leg Assembly (refer to
page 5).
MR
MRI Safety Checks
Ensure each component below has the 'MR
Safe' product sticker and engraving in the
below locations:
Armrest Wing:
Locking Leg Assembly:
Daggerboard:
Safety Clearance
refer to page 8
| adeptmedicaltraining.com
Patient Management
Ensure insulation protocol is followed according
to individual facility guidelines. This includes
ensuring sufficient insulation is placed to
eliminate skin to skin conductive loops and that
the patient's skin does not contact the bore of
the magnet, when using the device.
|
Disinfection
!
Warnings
Insufficient cleaning may compromise
disinfection process and lead to
transmission of pathogens.
Do not use any disinfectant products or
abrasive/corrosive agents which are not
on the approved cleaners list.
Always read manufacturer's instructions
and consult the manufacturers MSDS for
cleaning and disinfectant products.
Some disinfectants may cause slight
discolouration to the soft blue material
used on some components within the
product range. This will not affect the
strength and the product will remain fit
for purpose.
Limitations on Processing
Discontinue use if:
Any cracks or breakages are present.
|

Disinfection Instructions

Initial treatment at the point of use
It is important to clean the product once it
is removed from the packaging and after
each procedure. Ensure all areas including
joints, clips, sockets, brackets and levers are
thoroughly cleaned after use to remove all
contaminant build-up that may be present
following a procedure.
EN
Preparation before cleaning
Remove Overhead Arm Support from
beneath the mattress.
Remove Arm Straps from Armrest Wing.
Disassemble Overhead Arm Support
into Armrest Wing, Locking Leg
Assembly and Daggerboard.
Ensure the Locking Leg Assembly is fully
extended.
Cleaning: Manual
If any visible residue is present, rinse
under water, but avoid submerging the
device.
A soft bristle brush can be used.
Ensure the device is completely dried
prior to disinfection.
Disinfection
Refer to the Approved Cleaners List (following
page) for selection of an appropriate
disinfecting agent.
Armrest Wing
With a cloth dampened by an approved
disinfectant, wipe all surfaces, including
inside the Arm Strap slots.
Daggerboard
With a cloth dampened by an approved
disinfectant, wipe all surfaces, including
inside of the Locking Leg Assembly slot.
Locking Leg Assembly
With a cloth dampened by an approved
disinfectant, wipe all surfaces, especially
the blue lever.
Arm Strap
With a cloth dampened by an approved
disinfectant, wipe all surfaces.