CORNING Matribot Bioprinter Руководство по эксплуатации - Страница 6

Просмотреть онлайн или скачать pdf Руководство по эксплуатации для Лабораторное оборудование CORNING Matribot Bioprinter. CORNING Matribot Bioprinter 8 страниц. Demonstration equipment
Также для CORNING Matribot Bioprinter: Краткое руководство по эксплуатации (2 страниц)

CORNING Matribot Bioprinter Руководство по эксплуатации
4. Lift the printbed support located at the back
of the printer to raise the printbed slightly
higher than required to fit the lower inner
foam insert. Push the lower inner foam
insert from the front of the printer to slide it
under the printbed. Lower the printbed onto
the foam by gently pushing on the printbed
support located at the back of the printbed.
5. Confirm that the thermal insulator has been
removed from the cooling block to prevent
damaging the thermal insulator. Gently
move the printhead to the center of the
unit and slide the middle inner foam insert
from the front as shown. Align the lower
and middle foam inserts with each other by
adjusting the position of the printhead.
6. Make sure the syringe plunger is straight
so it is aligned with the slot in the top
foam and hold the printhead cable out of
the way, ensuring it is aligned with the
cable cutout in the back of the foam insert.
Insert the top inner foam insert from the
top of the unit as shown.
7. Place the Corning® Matribot® bioprinter inside the plastic bag, ensuring that the
bioprinter is completely covered.
5