NOTE!
Store the loose components of the FLUSH 2.0
system in the enclosed case. You will need them
again if the flushing system is to be used without the
recording unit.
Re-fit the FLUSH 2.0 system to the appropriate tap
adaptor.
CAUTION!
Please also refer to the operating instructions for the
FLUSH 2.0 system
7.2 Software
The "AS DataLog" software is required for programming the
AS DataLog and evaluating the data recorded. The files and
respective instructions can be found on the device in the
"Software" folder.
8

Troubleshooting

It is possible that the "AS DataLog" system may not start
correctly or is not recognised as a Windows drive. In this
case, it is necessary to reset the system.
Please proceed as follows:
1. Remove the lid of the battery compartment.
2. Briefly press the "RESET" button.
6
9

LED status signal

The "AS DataLog" recording system indicates the device
status via an integrated LED.
LED
LED off
LED flashes 4 times at
intervals of 2 sec.
LED flashes once at
intervals of 30 sec.
Status
1. Battery flat; battery
needs to be changed
2. New battery inserted
but no data recording
programmed yet
2 s
The "AS DataLog"
system is connected to
a PC; no measurements
are being recorded. As
soon as the "Setup"
option is clicked in the
evaluation software, the
LED goes off.
30 s
Data recording is active
(data is saved every 30
seconds); no measure
required.
7