CPS Products BLACKMAX MD100 Series Руководство пользователя - Страница 18
Просмотреть онлайн или скачать pdf Руководство пользователя для Цифровые коллекторы CPS Products BLACKMAX MD100 Series. CPS Products BLACKMAX MD100 Series 20 страниц. Digital 2 and 4 valves
Find Quality Products Online at:
APPENDIX C (MDXBK CHARGER ACCESSORY SPECIFICATIONS)
Operating Temperature Range
Operating and Storage Humidity
Storage Temperature
Power Source
Battery Type
Rechargeable Battery Cycle Life
APPENDIX D (BATTERY SELECTION AND REPLACEMENT)
For optimum life and economic value, CPS Products, Inc. recommends that MD50 and MD100
manifold series be powered by rechargeable NiMH (nickel metal hydride) batteries. CPS Products
offers an optional "Charger Kit" (model MDXBK) that comes with a power adapter and 4 NiMH
batteries. Do not connect the battery charger if there are no rechargeable batteries in the manifolds.
NOTE: Alkaline batteries should not, and cannot be charged in the BLACKMAX MD50 or MD100
Series. Battery charge status is shown by the battery indicator symbol in the upper left [FIG.1]
Main Screen and all other screens. NOTE: A slight increase in temperature near the charging
socket is normal. Remove battery door and install 4 AA size batteries, observing battery polarity.
When using rechargeable batteries for the first
time, turn manifold off and connect one end
of charger to Manifold 9V DC plug [FIG.28]
and the other end to an appropriate wall outlet
socket. Charge NiMH batteries for 2 hours. The
Battery Symbol on the display will blink during
charging. MD50 and MD100 manifold series
can be used while charging the batteries, but
charging time will be considerably extended.
The preferred way to use the Rapid Charging
Feature is to charge while the manifold is OFF.
APPENDIX E (CLEANING SENSOR IN MDXVG)
Sensor may become contaminated with dust, oil, etc. from pulling vacuums.
1. Never use a cotton swab or similar object to clean the sensor. This may damage the
sensor.
2. Open all knobs, valves, and cap all the ports except for the VAC port.
3. Invert the MDXVG and add about 1 tablespoon of 100% denatured alcohol into the VAC port
(see FIG 2) so as to flush out contaminants.
4. Gently shake the unit to produce a swirling effect of the alcohol inside the housing.
5. Turn right side up, open a port to pour out the rubbing alcohol.
6. Repeat the above procedure until the alcohol appears to be clean and clear of contaminants.
7. Check the gasket located inside the VAC port. If worn or deformed, replace with a new CPS
gasket (PN: HXG)
8. Open all ports to allow sensors to air dry (usually about 1 hour). DO NOT BLOW DRY.
18
GlobalTestSupply
www.
32°F to 122°F (0°C to 50°C)
0-95%
-40°F to 185°F (-40 to 85°C)
100-240VAC input, 9VDC 2.0 amp maximum output
AA size, NiMH rechargeable, approximate battery life is 20,000 to
50,000 hours*
1000 recharges *based on usage with or without the LCD backlight
energized.
FIG.28
Connection
for 9V DC
Charger
.com