3B Medical Lumin LM3000 Руководство по эксплуатации - Страница 9

Просмотреть онлайн или скачать pdf Руководство по эксплуатации для Средства личной гигиены 3B Medical Lumin LM3000. 3B Medical Lumin LM3000 14 страниц. Instructions for healthcare facilities: bioburden reduction uv system user's manual
Также для 3B Medical Lumin LM3000: Руководство пользователя (7 страниц)

3B Medical Lumin LM3000 Руководство по эксплуатации
handling contaminated, compatible N95 respirators.
2. Check date on paper bag to ensure it has been held for at least 5 days prior to bioburden
reduction.
3. Confirm the compatible N95 respirator is model 3M 1860 (aqua colored exterior).
4. Confirm the name on the compatible N95 respirator matches the name on the bag.
5. Discard bag in biohazard waste bin.
6. Visually inspect the compatible N95 respirator for:
a) Wear (if damaged, discard and inform user)
b) Soil (if soiled, discard and inform user)
c) Number of tick marks (if 4 marks, discard and inform user).
Important: If the compatible N95 respirator has signs of wear, soil, or has lost traceability (name or
number of cycles), discard the respirator and notify the user.
Bioburden reduction
1. Open Lumin LM3000 drawer with clean gloved hands.
Important: Lumin LM3000 works only when the compatible N95 respirator is directly exposed to the UV
light. Place a single compatible N95 respirator in the device.
2. Place compatible N95 respirator in Lumin LM3000 drawer, with the outer surface of the
compatible N95 respirator facing up. Please see graphic below.
3. Confirm that the elastic bands are not in the way of the UV light.
4. Change gloves.
5. Place an unused UV indicator strip, with sensitivity of 1 J/cm
acrylic base of the drawer next to the compatible N95 respirator. This will change color if a
complete dose has been delivered. 3B Medical recommends CONTROL-CURE® UV FASTCHECK™
STRIPS for 1,000 mJ/cm
6. Close the drawer and press button to initiate 5 minute cycle. The Lumin LM3000 is pre-
programmed for a 5 minute cycle.
Caution: Do NOT open the drawer before the cycle is completed. If the drawer is opened before the
cycle is complete, the compatible N95 respirator has not been fully irradiated. Repeat the complete 5
minute cycle.
7. Change gloves.
8. At conclusion of cycle, confirm the UV indicator strip has changed color to confirm a full dose of
UV has been delivered.
Caution: If the strip does not change color, the compatible N95 respirator has not been exposed to a
sufficient dose of UV and must be put aside. Call 3B Medical for guidance and possible service. Do not
use the device until 3B has provided instructions or serviced the device.
Important: While the compatible N95 respirator is being exposed to UV, the surface of the UV drawer
that is under the compatible N95 respirator is in contact with a contaminated respirator surface, and it is
not being irradiated with UV because it is shielded from UV by the respirator. Therefore, the bottom of
the drawer is contaminated and must be cleaned before the compatible N95 respirator is turned over, to
prevent re-contamination of the outer surface of the respirator, which was already exposed to UV.
9. With clean gloves, remove the compatible N95 respirator and place on clean paper towel, with the
outer surface of the compatible N95 respirator facing up.
10. Clean base of drawer with alcohol wipe or run a cycle of the Lumin LM3000 with no respirator to
remove pathogens that remain in the drawer of the Lumin LM3000.
11. Change gloves.
12. Place the compatible N95 respirator in drawer, with the inner surface of the compatible N95 respirator
facing up.
13. Confirm that the elastic bands are not in the way of the UV light.
2
exposures.
, in the drawer, facing up on the
2
9