Olympus CH20 Руководство по ремонту - Страница 16
Просмотреть онлайн или скачать pdf Руководство по ремонту для Микроскоп Olympus CH20. Olympus CH20 31 страница. Biological microscope
![Olympus CH20 Руководство по ремонту](/uploads/products/server2/4282441/4282441.jpg)
CH10/CH20
2. Revolving Nosepiece
No.
Parts name
•
SCREW
‚
NOSEPIECE
SPRING
ƒ
SHAFT
„
NOSEPIECE
...
RING
†
KNOB
C. DISASSEMBLY AND ASSEMBLY PROCEDURES
Screw
ANU5X6SA (*1)
B3SO (*2) 10 pcs.
B3SO (*3) 66 pcs.
Grease
Adhesive
OT2008
OT1028
OT2008
OT2008
C-6
Remarks
Absolute optical axis: within 0.3
mm
Click removing force: 6.4 - 8.8N
{650 - 900g}
Optical axis adjustment of
revolving nosepiece (See Repair
Procedure, D-1.)
Apply grease on the balls
(B3SO) fitting surface.
Apply adhesive on the screw
thread.
Apply grease on the balls
(B3SO) fitting groove.
Apply grease on the screw
thread.