D. REPAIR PROCEDURE
9-2
Adjusting parfocality in the right sleeve.
I
Work
I
I
Image seen through the standard eyepiece
I
1)
Loosen the
two
screws
that
fix the
I The cross hairs of the standard eyepiece are clearly
ADJ.RING 0 of the right sleeve.
I , visible, but the specimen in the standard objective is not
I clear.
I
I
I
I
I
I
I
I
I
I
GR\
1
Screw
1
I
I
I
I
12) Observing through the standard eyepiece,
I In the best position, fine stripes in the specimen
turn the ADJ. RING 0 until the specimen in I background look sharp.
the standard objective becomes clearly
I
I
visible.
I
I
(3) Tighten the screws that were loosened in step /
(1).
I
X The sleeve will ix deformed by forcefully
I
I
tightening the screws. After the screw end
I
contacts the sleeve, tighten the screw about 60"
1
to prevent deformation of the sleeve.
I
I
I
Make sure that the adjusted position is not displaced
tightening the screws.
D-24