D. REPAIR PROCEDURE
I
(5) Tighten the two screws that were loosened in
I
Make sure that the adjusted position is not displaced by
step (3).
( tightening the screws.
9-4
Check parfocality
Both the cross hairs of the standard eyepiece and the specimen in the standard objective must
look sharp. Even if any displacement is found, it should be within the following standard
range.
Standard range
The difference between the position where the cross hairs of the standard eyepiece look sharp and the position
where the specimen in the standard objective looks sharp should be within 2 graduations on the helicoid scale
of the KNO048.
Scale
If it is not within the standard range, return to "Adjusting parfocality" on page D-23 and
repeat the work.
D-26