Olympus MX61L Руководство по сборке - Страница 23

Просмотреть онлайн или скачать pdf Руководство по сборке для Микроскоп Olympus MX61L. Olympus MX61L 49 страниц. 200mm/300mm compatible semiconductor/fpd inspection microscopes

Olympus MX61L Руководство по сборке
Fig. 30
Setting the DIP Switches and
14
PRESET Switches
DIP Switch Setting (Fig. 30)
# Confirm the correct connections before proceeding.
}Set the main switch of the microscope to "
DIP switch @.
The DIP switch settings are read and entered at the moment the
microscope is turned ON.
# Be careful not touch the internal circuit boards when setting the DIP
switches. Otherwise, the internal circuitry may be damaged by static
electricity. Since the human body is charged with a small amount of
static electricity, discharge it from your body before proceeding to
the setup. This is done by lightly touching any metallic object.
Function
Revolving
nosepiece type
(Number of holes)
Aperture iris diaphragm
control button setting
Glare prevention during
objective switching
(Darkfield observation)
" (OFF) before setting the
Setting Result
U-D5BDREMC/P5REMC
OFF
U-D6REMC
ON
Setting enabled.
Setting disabled.
Glare prevention
No glare prevention
MX61/MX61L
(Fig. 30)
DIP Switches
1
2
3
4
OFF
ON
OFF
ON
: Factory setup.
20