Objective/Aperture Iris Switching
Functions
1. Rotates the objective in the specified revolving nosepiece position in the optical axis and also sets the aperture iris
diaphragm to the specified position.
2. When this command is sent during brightfield observation, the aperture iris diaphragm is set to the position specified with
the AS PRESET switch.
3. When this command is sent during darkfield observation, the aperture iris diaphragm is set to the open position.
4. When this command is sent during darkfield observation for which the glare prevention is set, the aperture iris diaphragm,
is closed, the revolving nosepiece is rotated and, after completing the rotation, the aperture iris diaphragm is opened.
Special notes
1. When this command is sent during brightfield observation, the revolving nosepiece and aperture iris diaphragm are
operated simultaneously, and the response is returned when both operations have completed.
2. If the request command for a 6-position is issued while a 5-position revolving nosepiece is used, an illegal parameter
error will occur.
3. Even when multiple errors occur, the error code for only one error is returned.
Command descriptions
Command
Type
1OBA
n
Request
(1 to 6 when a 6-
position nose-
piece is used, or
1 to 5 when a 5-
position nose-
piece is used)
*
)
"
" represents a space.
Line feed command <CR><LF> is required after commands and line feed command <CR><LF> are added after any
response.
Description
Response
Rotation of the revolv-
1OBA
ing nosepiece posi-
tion (objective) with
the number specified
1OBA
with "n" in the optical
axis, plus setting of
1OBA
the aperture iris dia-
phragm to the posi-
tion specified with the
AS PRESET switch.
1OBA
1OBA
1OBA
Result
+
Successful
The specified revolving nose-
piece position is engaged in the
optical axis.
!,E11214
Error
It cannot be engaged in the op-
tical axis due to a time-out error.
!,E01130
Error
The nosepiece is not connected
or the connected nosepiece is
not the designated model.
!,E11218
Error
An abnormality with the nose-
piece.
!,E11278
Error
The connection is disconnected
due to an abnormality with the
aperture iris diaphragm.
!,E01120
Error
A number or string other than "1"
to "6" is sent as "n".
MX61/MX61L
1OBA
Description