Accuris E2301 Series Руководство по эксплуатации - Страница 4

Просмотреть онлайн или скачать pdf Руководство по эксплуатации для Источник питания Accuris E2301 Series. Accuris E2301 Series 6 страниц. 300v power supply

2. Product Overview

In CONSTANT CURRENT mode, the power supply will output
4 Pairs of
the set mAmp current parameter, and the voltage will fluctuate
Touch Screen
Electrical Outputs
relative to the resistance of the load (buffer concentration in the
Control and Display
attached electrophoresis tanks) up to a maximum of the voltage
setting that has been chosen.
FRONT
6.4 Automatic Cross Over
During an electrophoresis run, the power supply will attempt to
maintain the constant voltage or amperage setting, depending on
which parameter is set to "CONSTANT".
If, during a run, the non-constant parameter setting (limit) is met
due to the resistance of the load, the power supply will
automatically "cross over" so the non-constant parameter will
now be the constant and limiting parameter. A red outline will
appear around the new "constant" value.
6.5 Maximum Power Limit
Folding
Recesses for
Front Feet
Stacking 2 Units
The myVolt Touch Power Supply has a maximum power output
of 100 Watts.
Per Ohms Law:
Power (P) = Current (I) x Voltage (V)
If the maximum power is reached during an electrophoresis run,
the non-constant parameter will be limited to prevent exceeding
.
the maximum power output
Cooling
CAUTION:
It is not recommended to run the power supply at or
Fan
near the maximum wattage, as this will cause the power supply
and gel tank to get hot. Most standard electrophoresis buffer
concentrations will not have such low resistance that the current
Rating, Serial #
AC Power Inlet
REAR
and voltage level will come close to the maximum power limit.
Label
and Fuse
7
2