Installation of the Leads and Current Sensors
Color-coded ID markers are supplied with the PowerPad
terminals.
• Detach the appropriate inserts from the color-coded marker and place them in the
holes provided under the terminals (larger inserts for current terminals, smaller inserts
for voltage terminals).
• Clip the rings of the same color to the ends of the lead that will connect to the terminal.
SD-Card
SD-Cards are supported. Note that the instrument's recording capacity is 2 GB maximum.
To Access the SD-Card:
• Make sure that the instrument is disconnected and off.
• Use a screwdriver or coin to unscrew the two screws of the battery compartment
cover.
• Remove the cover and withdraw the battery from its compartment without disconnect-
ing it.
• Press on the SD-Card to release it then press on the protecting tab to withdraw it from
its slot.
• When replacing the SD-Card, the contacts must be on the left side, and the locator
down.
• Slide it into its slot until it snaps into place. The protecting tab is at the top of the card.
• Put the battery back in its compartment and screw the cover back on.