Direct Healthcare Services Mercury Advance SMARTcare Руководство по эксплуатации - Страница 6
Просмотреть онлайн или скачать pdf Руководство по эксплуатации для Медицинское оборудование Direct Healthcare Services Mercury Advance SMARTcare. Direct Healthcare Services Mercury Advance SMARTcare 20 страниц. Pressure relieving mattress
3. Troubleshooting
Warning / Fault
Cause
Control Unit does
The Control
Unit may not be
not operate;
attached to a
no display lights
power source or
illuminate
a fuse may need
replacing.
Warning LED C
Mains failure /
Other
+ Audible Warning
See above,
plus:
Warning LED B
Pressure too low
+ Audible Warning
Warning LED B+C
Pressure too low /
Air pipe kinked
+ Audible Warning
Warning LED A
Pressure
too high
+ Audible Warning
Warning LED A+B
Alternating
Mode Failure (no
+ Audible Warning
alternation)
Warning LED A,B+C
Initialising Failure 1. Press the Audible Warning mute button to silence the Audible Warning.
+ Audible Warning
6
M E R C U R Y A D V A N C E S M A R T C A R E
Solution
1. Check the Control Unit is connected to mains power outlet with the correct voltage.
2. Check the Control Unit is switched on. Switch off and unplug the unit before restarting.
3. Check the mains plug fuse (5 AMP) then check both Control Unit fuses (1 AMP) –
fuses can be released using a screwdriver to push and turn.
Do not try to open the Control Unit. Opening the unit could cause personal injury
or equipment damage.
Ensure the replacement of fuses is carried out accordance with local legislation.
1. Check the Control Unit is connected to mains power outlet with the correct voltage.
2. Check the Control Unit is switched on. Switch off and unplug the unit before restarting.
3. Check the mains plug fuse (5 AMP) then check both Control Unit fuses (1 AMP) – fuses can be
released using a screwdriver to push and turn.
Do not try to open the Control Unit. Opening the unit could cause personal injury or equipment
damage.
Ensure the replacement of fuses is carried out accordance with local legislation.
1. Reset the warning – turn off power and press the Audible Warning mute button.
2. Check that the Dual Function Connector is firmly attached the Control Unit (located on the left
of the Control Unit case). Check all air hoses along the inside of the mattress – each should be
firmly connected. Check each air cells is securely attached to its connecting pipe.
3. Check all cells, pipes and hoses for any air leakage.
4. Check that the air filter cover is correctly secured and the air filter is clean.
5. Switch on power.
1. Check the blue external umbilical air pipe that is between the mattress and the Dual Function
Connector is not kinked, twisted or damaged and that the CPR Tag is firmly secured.
2. Check all air hoses along the inside of the mattress – each should be firmly connected.
3. Check each air cell is securely attached to its connecting air pipe.
1. Reset the warning – turn off power and press the Audible Warning button.
2. Disconnect the air hoses to reduce pressure, reconnect when pressure has decreased.
3. Check for twists in the air hoses between Mattress and Control Unit.
1. Reset the warning – turn off Power and press the Audible Warning mute button.
2. Disconnect the air hoses to reduce pressure – reconnect when pressure has decreased.
2. Check all air hoses along the inside of the mattress – each should be firmly connected.
Check each air cell is securely attached to its connecting air pipe.
3. Check that the Dual Function Connector is firmly attached the Control Unit (located on the left
of the Control Unit casing).
D I R E C T H E A L T H C A R E G R O U P . C O M