Pressure management:
This replacement mattress system is designed as a tool to help you to provide
pressure management in the prevention and treatment of pressure ulcers.
Operating instructions:
The Panacea Air Pro System incorporates alternating pressure therapy techniques
to reduce interface pressures on the skin surface. When the resident is placed on
the mattress surface, the mattress firmness is adjusted automatically after the
resident's height and weight are entered on the display panel. This distributes the
body weight uniformly across the mattress surface. The Alternating Pressure mode
reduces the air pressure in alternate air cells every 5, 10 or 15 minutes to reduce
the interface pressure on specific areas of the body even further. The air cells with
reduced air pressure are alternated with adjacent air cells to allow a major portion
of the body to obtain the therapeutic effect while maintaining the resident in an air
flotation state.
Installation of the Panacea Air Pro System
The Panacea Air Pro Mattress can be placed on any conventional bed used in
hospitals, nursing homes or a home healthcare bed. The original bed mattress
should be removed and stored in an appropriate place. The following steps should
be completed in installing the system:
1. Roll out mattress onto the bed frame with the air hoses exiting at the foot of the bed.
2. Strap the mattress base to the bed frame. Use caution when attaching the straps to
make sure there is no strap damage when the bed position is changed.
3. Place the Panacea Air Pro System Controller on the foot board as close to the center
as possible.
4. Make sure the hoses can reach the controller without undue strain.
5. Connect the air hoses to the controller.
6. Press the hose onto the controller connector until the hose connector clicks.
7. Pull back gently on the hose connector to verify the mechanical connection is secure.
8. The appropriate connections will be obvious as only certain mattress side connectors
will mate properly with the control unit side ports.
9. Examine each of the air cells in the mattress. Make sure each air cell is vertical and
not twisted in the mattress base.
10. Verify positive air connections by gently pulling the air manifold and air cell
connections gently apart.
11. If the connectors separate, press them back together until a click is heard.
12. Loose air cell connectors may prevent the mattress from inflating properly.
13. Check the Velcro tabs on the end of each air cell to make sure they are securely
fastened to the mattress base.
14. Route the power cord in a safe manner to the closest power outlet and plug in the unit.
15. When the unit is powered up, set the control unit mode to Max Inflate mode. This
will fill the mattress more rapidly in preparation for the resident.
6

Product Description

(cont)