16. Note the inflation of the air cells. Make sure all air cells are standing up straight and
are not twisted in the mattress base.
17. Change the mode from Auto-Firm mode to the desired mode of operation.
18. Adjust the firmness level for the resident by pressing the Increase/Decrease
Firmness Button. Weights on label are for reference only and are not a substitute for
a physical hand check.
SYSTEM FUNCTIONS
Power Button
Toggles between Power On and Standby mode.
• Standby mode: Amber LED on, main power is on but pump is off
• Power On mode: Green LED on, pump is operating
• If no mode is selected, the pump defaults to Static Pressure mode
• Hold the power button down for more than 2 seconds to enter Standby mode
Modes of Therapy Button
Toggles between Static Pressure mode and Alternating Pressure modes programmed
into the control unit. The cycle time settings for Alternating Pressure mode are 5-, 10-,
15- and 20-minute periods.
Low Air Loss
Provides airflow to top cover of mattress. Available only in Static Pressure mode.
Auto-Firm Mode
Inflates mattress to maximum firmness.
Alarm Silence Button
This turns off the audible alarm (the LED for the warning still flashes). If the alarm
condition is re-triggered, the audible alarm will be turned on again.
— The Panacea Air Pro top cover is designed for use without a top sheet. If a bottom
sheet is desired, place a flat sheet over the top cover and allow it to drape over on
all four sides of the bed. Do not tuck in the ends of the sheet under the mattress.
The bottom sheet must have the ability to conform to the resident's body profile.
This relieves excess pressure and shear on the skin.
• Power Failure Indicator/Alarm – This LED blinks and sounds an audible alarm to
indicate power has been removed, either because the plug was removed or power to
the outlet has been removed.
• Low Pressure Warning – If a problem exists with maintaining pressure in the active
alternation channel, this LED will blink and an audible alarm will sound.
Lock Button
When the lock button is depressed for 2 seconds this button will disable all of other
buttons on the controller to prevent unauthorized or accidental changes from occurring.
This function can be disabled by depressing the button for an additional 2 seconds. The
audible alarm is sounded to indicate the operation of this function and a yellow LED
indicates the locked state.