Installation and Setup
1. Remove the system from the shipping container and place the unit on the lab
bench.
2. Remove all packing material used to ship the unit (i.e. packing inserts, tie
wraps, etc...).
3. Connect the AC line cords to a stable power source.
Install one-liter glass beakers in all positions. Add 900 mL. of media to each
4.
beaker.
5. Attach the disintegration baskets to each arm.
6. Turn on the system power using the switch on the side of the unit.
7. After the system goes through a self-test, it will enter Scan Mode and the
'Station No.' will scroll through temperatures for each stations.
Setting Beaker Temperature and Test Time for Individual Positions
1. If in Scan Mode press the up or down arrow keys to exit this mode.
2. Scroll to the desired basket position using the up or down arrow keys.
3. To set beaker temperature press the PRGM button. The temperature set point
will flash.
4. Raise or lower the set temperature using the up or down arrow keys.
5. Press the TIME button.
6. Raise or lower the test time using the up or down arrow keys.
7. Press ENTER to accept the values.
Setting Beaker Temperature and Test Time for All Positions
1. If in Scan Mode press the up or down arrow keys to exit this mode.
2. Press the PRGM button twice. Station number '0' will be displayed and the
temperature set point will flash.
3. Raise or lower the set temperature using the up or down arrow keys.
4. Press the TIME button.
5. Raise or lower the test time using the up or down arrow keys.
6. Press ENTER to accept the values.
Enabling All Positions
1. Press the CLEAR/SCAN button to enter Scan Mode.
2. To activate or deactivate all positions press the ENABLE/DISABLE button.
Observe the ENABLE LED illuminate next to the station number on the display.
DISTEK, INC.
Quick Start Instructions: Bathless
Disintegration System
Quick Start Instructions
Bathless Disintegration System
Rev: B
Doc: 3800-0447
Rev Date: 12/02/03
Doc. No. 3800-0447
By: JSB
Page 1 of 2