Quick Start Instructions
Bathless Disintegration System
Starting Preheat
1. To preheat a particular beaker select the station number and press START.
2. To preheat all enabled beakers press START ALL.
Beginning a Test
1. Select which station to begin by setting this station as the active position using
the arrow keys.
2. Press the RAISE button to raise the beaker.
3. Introduce tablets within 2 minutes.
4. Press the LOWER button to lower the beaker and begin the test.
5. Repeat steps 1-4 for the remaining positions.
Viewing Information during a Test
1. While a test is running set the unit to Scan Mode (pressing the CLEAR/SCAN
button will enable the instrument to scan through all station numbers) or
Individual Position Mode (using the arrow key to view data on specific stations
only).
2. Press the T°C button to display the current temperature.
3. Press the TIME button to display the test time information. Pressing the TIME
button repeatedly will toggle between elapsed and remaining time (notice the
ELAPSED and REMAINING LEDs next to the time illuminate).
4. Press the STATUS button to view the test status. The options are Preheat,
Ready, Running and Idle.
Aborting a Test
Aborting All Stations:
1. While a test is running press STOP ALL. The screen will flash 'ABORT?'.
2. Press ENTER to abort or CLEAR/SCAN to return to the test (cancel).
Aborting Individual Stations:
1. While a test is running use the arrow keys to select the station to abort.
2. Press ENABLE/DISABLE. The screen will flash 'ABORT?'.
3. Press ENTER to abort or CLEAR/SCAN to return to the test (cancel).
DISTEK, INC.
Quick Start Instructions: Bathless
Disintegration System
Rev: B
Doc: 3800-0447
Rev Date: 12/02/03
(continued)
By: JSB
Page 2 of 2