Maintenance log L2-Range and D2-Range
With Combi-Sepamatic CS1: Perform a visual leak test. Check function and suction
power: Observe the manufacturer's (Dürr) maintenance instructions.
❯ Cleaning of the suction unit in accordance with the operating instructions. ❯ Clean
or replace the protective sieves at the aspiration inlet. ❯ If a rinsing unit is present:
clean the sieve in the water supply. ❯ Perform a functional test.
With Combi-Sepamatic CS1: Perform a visual leak test. Check function and suction
power: Observe the manufacturer's (Dürr) maintenance instructions.
❯ Replace the rubber grommets on the connections.
With Combi-Sepamatic CS1: Perform a visual leak test. Check function and suction
power: Observe the manufacturer's (Dürr) maintenance instructions.
❯ Replace the rubber grommets on the connections.
❯ Replace all O-rings in the device.
Visual leak test.
9. Suction
Suction filter block check suction filter for soiling and proper drain; clean it if
necessary.
Suction filter block: check sealing cap of the suction filter drawer and replace
O-ring.
Suction filter block: check filter sieve drawer for defects and replace if necessary.
Suction filter block: pull off suction hoses and replace O-rings.
Check suction hoses for defects and replace if necessary.
Suction shut-off valve: Perform a visual leak test. Clean and replace valve
membrane of the valve.
Visual leak test.
L2-Range
Once a year
D2-Range
Every 3
L2-Range
Replacement parts set
years
D2-Range
Every 5
L2-Range
years
D2-Range
L2-Range
Inspection
D2-Range
L2-Range
Once a year
D2-Range
L2-Range
Once a year
O-ring seal DIN3770-40x2,5
D2-Range
L2-Range
Once a year
D2-Range
O-ring seal DIN3770-20x2
L2-Range
Once a year
D2-Range
O-ring seal DIN3770-14x2 (2
L2-Range
Once a year
D2-Range
L2-Range
Membrane for suction air
Once a year
D2-Range
L2-Range
Inspection
D2-Range
(order from
-
DÜRR Dental)
7117-980-33
(order from
(3 years)
DÜRR Dental)
7117-980-22
O-ring set for CS 1
(order from
DÜRR Dental)
-
-
-
-
640000
(included in
maintenance
set 211390)
514100
(included in
Filter for suction bloc
maintenance
set 211390)
640001
(included in
maintenance
set 211390)
640002
pcs)
(included in
maintenance
set 211390)
-
-
513500-1
(included in
shut-off valve
maintenance
set 211390)
-
-
-
-
-
-
-
1
1
1
2
-
1
-
page 8