ARTRAY ARTCAM-036MI2 Series Буклет с инструкциями - Страница 19
Просмотреть онлайн или скачать pdf Буклет с инструкциями для Цифровая камера ARTRAY ARTCAM-036MI2 Series. ARTRAY ARTCAM-036MI2 Series 20 страниц. Usb2.0 cmos camera
8-2-3 Trigger timing chart (ARTCAM-500MI series)
Trigger Input
TRIG Signal
Reset Time
Exposure Time
st
1
Line
nd
2
Line
Last Line
Vertical Synchronization Signal
Image Transfer (USB)
t0
Input Pulse Width
t1
Photo Coupler Delay
t2
Reset Time
t3
Exposure Time
t4
Frame Transfer
t5
Effective Trigger for next frame
*For ARTCAM-500MI, the image will be acquired when the trigger signal transits from a low voltage to a high
one. If the trigger signal stays at a high level, the trigger time cycle will repeat automatically and image will be
acquired continuously.
t0
t1
t2
t3
t5
t4
Above 1ms
1~99μs (Subject to Input Pulse Voltage)
8 * 1H
Subject to Shutter Speed Settings
Subject to ROI Settings and Operation Clock
Approximate 100H after the Vertical Synchronization Signal
transits to a low voltage
- 19 -