3. INSTALLATION INSTRUCTIONS
3.1
General
The ORTEC 416, used in conjunction with the ORTEC 401A/402A Bin and Power
Supply, is intended for rock mounting; therefore, it is necessary to ensure
that vacuum tube equipment operating in the same rack has sufficient cooling
air circulating to prevent any localized heating of the oM-tronsistor circuitry
used throughout the 416. The temperature of equipment mounted in rocks can
easily exceed the recommended maximum unless precautions ore token. The
416 should not be subjected to temperatures in excess of 120°F (50®C).
3.2 Connection to Power — AEC Standard Nuclear Instrument Module, ORTEC
401iV402A
The 416 contains no internal power supply; therefore, it must obtain power
from a Nuclear Standard Bin and Power Supply such as the 401A/402A. It is
recommended that the Bin power supply be turned off when inserting or re
moving modules. The ORTEC 400 Series is designed so that it is not possible
to overload the Bin power supply with a full complement of modules in the
Bin; however, this may not be true when the Bin contains modules other than
those of ORTEC design. In this cose, power supply voltages should be
checked after the insertion of modules. The 401A/402A has test points on the
power supply control panel to monitor the dc voltages.