AMC 350 Talimatlar Kılavuzu - Sayfa 13

Verici AMC 350 için çevrimiçi göz atın veya pdf Talimatlar Kılavuzu indirin. AMC 350 17 sayfaları.

5 OPERATION AND CALIBRATION

5.1 OPERATION

The AMC-350 series sensor/transmitter unit is factory calibrated for the gas listed in Product
Information (section 2). The unit should not need recalibration when first installed and powered
up, but a test for correct operation is recommended.
stabilization period.
In general, after the stabilization period, the transmitter should be sending (in a Zero gas
environment) a signal of approximately 4 mA to the monitor or controller. However, there are a
few situations where a slightly higher or lower than normal signal may be noticed. In many
facilities there can be residual levels of the gas being detected in the air at all times. These can
cause a minor response from the sensor, normally causing a rise in signal. Other causes for
minor signal variations include extremes in temperature. In the case of large signal variations
(in a clean environment), check for an installation problem.

5.2 CALIBRATION

The sensor/transmitter is equipped with a remote calibration feature allowing one-man
calibration at the transmitter location. The transmitter output is measured using a plug-in type
"Remote Calibration Lead" designed to be adaptable to most multimeters. Zero and span
adjustments are made at the transmitter. Recalibration is necessary when replacing the sensor.
Verification of calibration should be done at least once every 6 months for safety reasons and
for highly demanding applications, monthly verification is recommended.
Factory/on-site calibration services, customer training and/or calibration adapter are available.
Specify the sensor/transmitter Part Number & Serial Number listed in section 2.1
SENSOR/TRANSMITTER and refer to section 2.2 FACTORY CALIBRATION for Gas Type and
calibration adapter when ordering any of the above.
Caution:
- Only qualified personnel should perform the actual calibration.
- Users are advised to consult The Armstrong Monitoring Corporation as to the
calibration procedure and recommended gas concentration for the application.

5.2.1 EQUIPMENT REQUIRED

Digital multimeter
Remote Calibration Lead (provided with the transmitter)
Trimmer adjustment tool or miniature screwdriver
Calibration adapter
Zero & Span gases and Permeation tube (Contact factory for information)
AMC-350 Sensor/Transmitter
11
All testing should be done after a