blinking on the display (Fig.16). Press "▲"
or "MIN/MAX▼" to change to "Yes", then press 
"SET " to save the setting. Press "ESC" to return 
to P4.
C without saving. 
0
 
 
 
 
 
 
 
If you choose "Yes", the meter is reset to the 
following factory defaults: 
 
 
 
 
 
 
P5.
 SET SAMPLING RATE 
0
Go into P5.0 to set the data logging sampling 
rate (Fig.17). The range is from 1 second to 4 
hours 59 minutes and 59 seconds.   
 
Press "SET" to see P5.1 blinking hour digits on 
the lower left display . To change the digit, press 
"▲" or "MIN/MAX▼" to adjust the hour 
sampling rate. Press "SET" to save the setting.   
 
After the hours are set, you will see P5.2 and 
the minutes digits will flash. To change the 
digits, press "▲" or "MIN/MAX▼" Press "SET" to 
save the setting and go to P5.3. 
 
After the minutes are set, you will see P5.3 and 
the seconds digits will flash (Fig.18). To change 
the digits, press "▲" or "MIN/MAX▼" Press 
"SET" to save the setting and return to P5.0. 
 
Press "ESC" at any time to exit without saving 
your last changes and return to P5.0. 
 
 
CO2 CALIBRATION 
The meter is calibrated at 400 ppm (parts‐ per‐
million) CO2 concentration at the factory. 400 
ppm is comparable to fresh outdoor air.   
There are 2 types of calibration available: ABC 
(Automatic Baseline Calibration) or manual 
calibration. One or the other should be used 
regularly to maintain good accuracy. 
 
 
 
Note: 
If TIM10 accuracy becomes a concern after long 
use or other special conditions, return the 
meter to the dealer for calibration. 
 
CAUTION: 
Do not calibrate the meter with air or bottled 
gas that contains an unknown CO2 level. The 
meter will assume it is 400 ppm CO2 and will 
give inaccurate measurements. 
 
ABC Calibration 
ABC (Automatic Baseline Calibration) is 
designed to eliminate the zero drift over time of 
the infrared CO2 sensor. The ABC function is 
enabled by default from the factory. 
 
For ABC to work properly, the meter must 
continuously monitor CO2 levels for at least 7.5 
days (power on). During that time, it will record 
the lowest CO2 level and set that as the 400 
ppm baseline for future readings. Therefore, it 
is critical when using ABC calibration that the 
meter be exposed to fresh air at least once 
every 7.5 days. If this is not possible, then 
manual calibration is recommended. 
 
Manual Calibration 
Manual calibration requires fresh, dry outdoor 
air when CO2 levels are approximately 400 
ppm. Do not calibrate during rainy days. High 
humidity will affect the CO2 levels in air.   
 
During manual calibration, avoid placing the 
meter in areas crowded with people or close to 
structures that might give off CO2 like air 
ventilation pipes, chimneys, or automobiles. 
 
To manually calibrate the meter, place it 
outdoors, and turn it on for at least 2 minutes. 
Then hold down "SET" ,"▲" and "MIN/MAX▼" 
simultaneously more than 1 second to enter 
CO2 calibration mode. 400 ppm and "CO2" will 
blink during calibration. 
 
 
 
 
 
 
Wait about 30 minutes until the blinking stops. 
Calibration is completed automatically. To stop 
manual calibration, press "RESET" for at least 1 
second. 
 
Note: 
Keep the meter away from any animals, 
humans or plants which might affect the CO2 
concentration nearby during the calibration. 
 
RH CALIBRATION 
This humidity accuracy of this desktop CO2 
meter is set at the factory, and cannot be re‐
calibrated. 
 
 
TROUBLESHOOTING 
Cannot power on 
      Check whether the adaptor has power and 
      that the meter is firmly plugged in. 
Slow response 
      Check that the air flow channels on 
      the rear of the meter are not blocked. 
No change in readings 
      Check whether the meter is in maximum 
      or minimum mode. Reset factory defaults. 
"BAt" and green LED keep flashing 
      The adaptor output voltage is too high or 
        too low. Use the supplied adaptor with   
        correct 5V (+10%) >=0.5A ratings. 
 
Error Codes 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 