Ameda Elite Talimatlar Kılavuzu
Set Up and Assembly
Ameda HygieniKit Milk Collection System
Before Every Use
Always wash hands well with soap and water before
handling the Ameda Elite Breast Pump and Ameda
HygieniKit Milk Collection System.
IMPORTANT NOTE: ONLY USE THE AMEDA HYGIENIKIT
MILK COLLECTION SYSTEM ON AMEDA ELECTRIC BREAST
PUMPS.
Prior to starting pump, assemble and connect the Ameda
HygieniKit Milk Collection System, per included instructions. If
the Ameda HygieniKit Milk Collection System is sterile, use as
is. If not sterile, see the Ameda HygieniKit Milk Collection
System cleaning instructions.
Adapter Cap
2
Silicone Diaphragm
1
Breast Flange
3
Valve
4
Milk Bottle
5
Tubing
Tubing Adapter
The Ameda HygieniKit Milk Collection System can be used
to single or dual pump.
To dual pump:
To single pump:
Remove one tube and
Both tubes should be
close tubing adapter.
attached to tubing
adapter.
Elite with Detachable Power
Cord
• Attach Ameda supplied
power cord to AC power
inlet on rear of pump.
• Connect power cord to wall
outlet.
Elite with Built-in Power
Cord
• Connect power cord to wall
outlet.
6
E N G L I S H
• Insert tubing adapter into
tubing adapter port on
piston/cylinder.
Operating the Pump
• Center nipple(s) in breast
flange(s) and fill breast
flange(s) with the breast(s)
to create an air seal.
• Turn the pump ON by turning
the 'VACUUM' dial to the
right.
Setting Suction
• Increase the suction by
turning the 'VACUUM' dial to
the right. Increase suction
until reaching maximum
comfort level.
• If discomfort is felt in nipple(s)
or breast(s), decrease suction.
If discomfort is felt at all
suction settings, turn pump
off, insert a finger between breast(s) and breast flange(s)
to break suction, remove breast flange(s) from breast(s).
Discontinue pumping and contact a healthcare provider.
Setting Speed
• Start with the 'CYCLES' dial
all the way to the right to
increase the speed and
trigger a milk ejection reflex
(let-down)
• When milk flows, (let-down)
turn the 'CYCLES' dial to the
left to slow down the speed
to help drain the breast
faster. When milk flow slows, turn the 'CYCLES' dial all the
way to the right to help start another (let-down). Repeat as
milk flow changes to pump more milk.
Turning Pump OFF
• When finished pumping, turn
the 'VACUUM' dial all the
way to the left to turn the
pump off.
Removing Ameda HygieniKit Milk Collection System
from Breast(s)
• After pump is switched off, if necessary, insert a finger
between breast(s) and breast flange(s) to break suction.
• Remove breast flange(s) from breast(s).
Cleaning the Ameda Elite Breast Pump
• Disconnect the pump
power cord from wall
outlet.
• Remove piston cylinder by
placing index finger into the
tube, lifting upwards and
turning right to release
from attachment bars.
For At-Home/End-Users
After each pumping, wipe
pump with cloth dampened
with warm, soapy water.
For Hospitals/Rental Stations
Clean between users.
Spray with recommended
disinfectant, such as
CaviCide
TM
. Wipe with soft
cloth. Do not use products
containing phenol.
Do not use bleach on pump.
• Place piston cylinder back
onto attachment bars and
press gently into place with
thumb.
Troubleshooting
Pump Does Not Turn ON
• Ensure proper power cord connection to pump.
• Ensure power cord connection to wall outlet.
• Ensure outlet has power.
Low or No Suction
• Check breast flange for adequate seal against breast.
• Check that the Ameda HygieniKit Milk Collection System
is properly assembled.
• Check that adapter cap is firmly snapped onto top of
breast flange.
• Ensure silicone diaphragm is in place inside top of breast
flange.
• Ensure tubing adapter is properly fitted into tubing
adapter port on pump.
• Inspect valve for stretching or tears. If seen, replace
valve.
• Ensure valve is pushed firmly into place in lower portion
of breast flange.
• Ensure one side of tubing adapter is closed when single
pumping.

Maintenance

It is recommended the Ameda Elite Breast Pump be
inspected at least once every year to include the following:
• Visually check pump housing for cracks, breakage, or
damage and ensure pump feet are present.
• Visually check for power cord damage.
• Check for proper operation of controls and buttons.
• Check there are no missing screws.
• Check tubing adapter port on piston/cylinder is free from
obstruction.
Quick Functional Check
The following checks/tests should be performed prior to use
by a new user:
• Visual checks for power cord damage.
• Check for abnormal noise.
• Check for proper operation of controls.
Replacing Detachable Power Cord (if applicable)
• Power cord is detachable from power inlet.
• If power cord condition is in doubt, only replace with
Ameda supplied 2 conductor cord.
Technical Information
Main voltage
Elite with Detachable Power Cord 100VAC to 240VAC
Elite with Built-In Power Cord 100/120V ~ or 220/240V ~ /AC
Frequency
50 / 60 Hz
Power consumption
30 VA
Main fuse
T 1.6A
Protection class
Class II
Double isolated
Safety level
= BF
Drip protection
IP 21
Operational conditions +10 - +40°C
10 - 90% rel. humidity
Transport and
- 40 - +70°C
Storage conditions
10 -100% rel. humidity
Weight
6.4 lbs/2.9 kg
Dimensions
10.0"L x 9.0"W x 7.5"H
22.2cmL x 26.7cmW x 18.4cmH
Vacuum max.
[mbar]
[mm Hg]
Single HygieniKit
350 ( -10% )
262 ( -10% )
Dual HygieniKit
310 ( -15% )
232 ( -15% )
Variable cycle
30 - 60 / Min
Conforming to
0086
E325189 MEDICAL EQUIPMENT
WITH RESPECT TO ELECTRIC SHOCK,
FIRE AND MECHANICAL HAZARDS ONLY.
IN ACCORDANCE WITH
IEC60601-1: 1988 + A1:1991 + A2:1995,
IEC60601-1-2, CAN/CSA C22.2 No. 601.1-M90
Evenflo Company, Inc.
1801 Commerce Drive
Piqua, Ohio 45356
USA
EMERGO EUROPE
Molenstraat 15
2513 BH, The Hague
The Netherlands
Phone: +31.70.345.8570
Fax:
+31.70.346.7299
E N G L I S H
7