Use the Pressure Standard to output 500.00 psi (100%). After the
output has stabilized, press the ENTER key to continue. As the unit
takes readings, the screen will show -------. When the readings are
complete the screen should look as shown in the illustration that follows.
Use the Pressure Standard to output 437.50 psi (87.5%). After the
output has stabilized, press the ENTER key to continue. As the unit
takes readings, the screen will show -------. When the readings are
complete the screen should look as shown below.
Use the Pressure Standard to output 375.00 psi (75%). After the output
has stabilized, press the ENTER key to continue. As the unit takes
readings, the screen will show -------. When the readings are complete
the screen should look as shown below.
437.50
CAL MODE
375.00
CAL MODE
312.50
CAL MODE
PSI
PSI
PSI